Purpose:Optical coherence tomography angiography (OCTA) allows delineating the foveal avascular zone (FAZ) easily and noninvasively. The present study aims to test reproducibility and reliability of FAZ evaluation by means of OCTA in different settings. Methods:Twenty-four eyes of 24 normal subjects were investigated using AngioVue OCTA Imaging System. A series of OCTA acquisitions were taken both in basal and in different experimental settings after vasoactive stimuli. Images were evaluated separately by two operators and FAZ area was measured both manually and using the built-in automated measurement tool. Results:No differences for FAZ area were found in the repetition of basal acquisitions, neither in manual nor in automated measurement (0.215 ± 0.06 vs. 0.216 ± 0.07, and 0.268 ± 0.05 vs. 0.264 ± 0.09, first vs. second basal measurement in square millimetres for manual and automated evaluation, P = 0.25 and P = 0.35, respectively). Interoperators correlation was optimal (r2 = 0.978 [95% CI 0.981-0.976]). No differences were found among the other settings, which included first basal and then repeated (second) in the morning, after flickering light stimulus, after a Bruce treadmill stress test, after 30 minutes dark adaptation, and basal in the evening, neither in automated nor in manual measurements. Automated measurements for nonflow areas provided significantly larger diameters than manual ones. Conclusion:AngioVue OCTA Imaging System produces highly reproducible FAZ images with a high interoperators concordance level. Optical coherence tomography angiography capability to detect FAZ area seems not to be influenced by any of the vasoactive stimuli considered in the current study. Nonflow areas seem to be larger when measured automatically than manually.

Reproducibility and reliability of optical coherence tomography angiography for foveal avascular zone evaluation and measurement in different settings / Spina, C.L., Carnevali, A., Marchese, A., Querques, G., Bandello, F.. - In: RETINA. - ISSN 0275-004X. - 37:9(2017), pp. 1636-1641. [10.1097/IAE.0000000000001426]

Reproducibility and reliability of optical coherence tomography angiography for foveal avascular zone evaluation and measurement in different settings

Querques G.;
2017

Abstract

Purpose:Optical coherence tomography angiography (OCTA) allows delineating the foveal avascular zone (FAZ) easily and noninvasively. The present study aims to test reproducibility and reliability of FAZ evaluation by means of OCTA in different settings. Methods:Twenty-four eyes of 24 normal subjects were investigated using AngioVue OCTA Imaging System. A series of OCTA acquisitions were taken both in basal and in different experimental settings after vasoactive stimuli. Images were evaluated separately by two operators and FAZ area was measured both manually and using the built-in automated measurement tool. Results:No differences for FAZ area were found in the repetition of basal acquisitions, neither in manual nor in automated measurement (0.215 ± 0.06 vs. 0.216 ± 0.07, and 0.268 ± 0.05 vs. 0.264 ± 0.09, first vs. second basal measurement in square millimetres for manual and automated evaluation, P = 0.25 and P = 0.35, respectively). Interoperators correlation was optimal (r2 = 0.978 [95% CI 0.981-0.976]). No differences were found among the other settings, which included first basal and then repeated (second) in the morning, after flickering light stimulus, after a Bruce treadmill stress test, after 30 minutes dark adaptation, and basal in the evening, neither in automated nor in manual measurements. Automated measurements for nonflow areas provided significantly larger diameters than manual ones. Conclusion:AngioVue OCTA Imaging System produces highly reproducible FAZ images with a high interoperators concordance level. Optical coherence tomography angiography capability to detect FAZ area seems not to be influenced by any of the vasoactive stimuli considered in the current study. Nonflow areas seem to be larger when measured automatically than manually.
2017
37
9
1636
1641
Reproducibility and reliability of optical coherence tomography angiography for foveal avascular zone evaluation and measurement in different settings / Spina, C.L., Carnevali, A., Marchese, A., Querques, G., Bandello, F.. - In: RETINA. - ISSN 0275-004X. - 37:9(2017), pp. 1636-1641. [10.1097/IAE.0000000000001426]
Spina, C. L.; Carnevali, A.; Marchese, A.; Querques, G.; Bandello, F.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1404710
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 63
  • ???jsp.display-item.citation.isi??? 57
social impact