In this prospective, randomized, controlled study, we compared the performance of conventional and navigated total knee arthroplasties. Component alignment was measured in 60 patients operated on using navigation and in 60 patients operated on using the conventional technique. The groups then were divided into a subpopulation to measure alignments of the distal femoral cuts in the three anatomic planes, the proximal tibial cut in the frontal and sagittal planes, and the resulting lower limb mechanical axis in the frontal plane. Postoperative weightbearing long-view radiographs were evaluated as were clinical results using three standard questionnaires at 28 months followup. The intraoperative measurements (mean +/- standard deviation) at the resection planes showed navigated surgeries result in more accurate alignments than conventional surgeries for the femur: in the frontal plane, 0.1 degrees +/- 0.9 degrees and 0.7 degrees +/- 1.6 degrees valgus, respectively; in the sagittal plane, 1.1 degrees +/- 1.8 degrees and 2.8 degrees +/- 2.0 degrees flexion; and in the transversal plane, 0.1 degrees +/- 1.2 degrees and 0.9 degrees +/- 1.7 degrees internal rotation. The navigated technique also reduced the number of cases with final mechanical axes greater than 3 degrees from 20.0\% to 1.7\%. Postoperative radiographs showed better component alignment using navigation, particularly at the femur. However, clinical scoring systems showed this radiographic improvement did not necessarily result in a better clinical outcome at short-term followup.

Alignments and clinical results in conventional and navigated total knee arthroplasty / A., Ensini; Catani, Fabio; A., Leardini; M., Romagnoli; S., Giannini. - In: CLINICAL ORTHOPAEDICS AND RELATED RESEARCH. - ISSN 0009-921X. - ELETTRONICO. - 457:457(2007), pp. 156-162. [10.1097/BLO.0b013e3180316c92]

Alignments and clinical results in conventional and navigated total knee arthroplasty.

CATANI, Fabio;
2007

Abstract

In this prospective, randomized, controlled study, we compared the performance of conventional and navigated total knee arthroplasties. Component alignment was measured in 60 patients operated on using navigation and in 60 patients operated on using the conventional technique. The groups then were divided into a subpopulation to measure alignments of the distal femoral cuts in the three anatomic planes, the proximal tibial cut in the frontal and sagittal planes, and the resulting lower limb mechanical axis in the frontal plane. Postoperative weightbearing long-view radiographs were evaluated as were clinical results using three standard questionnaires at 28 months followup. The intraoperative measurements (mean +/- standard deviation) at the resection planes showed navigated surgeries result in more accurate alignments than conventional surgeries for the femur: in the frontal plane, 0.1 degrees +/- 0.9 degrees and 0.7 degrees +/- 1.6 degrees valgus, respectively; in the sagittal plane, 1.1 degrees +/- 1.8 degrees and 2.8 degrees +/- 2.0 degrees flexion; and in the transversal plane, 0.1 degrees +/- 1.2 degrees and 0.9 degrees +/- 1.7 degrees internal rotation. The navigated technique also reduced the number of cases with final mechanical axes greater than 3 degrees from 20.0\% to 1.7\%. Postoperative radiographs showed better component alignment using navigation, particularly at the femur. However, clinical scoring systems showed this radiographic improvement did not necessarily result in a better clinical outcome at short-term followup.
2007
457
457
156
162
Alignments and clinical results in conventional and navigated total knee arthroplasty / A., Ensini; Catani, Fabio; A., Leardini; M., Romagnoli; S., Giannini. - In: CLINICAL ORTHOPAEDICS AND RELATED RESEARCH. - ISSN 0009-921X. - ELETTRONICO. - 457:457(2007), pp. 156-162. [10.1097/BLO.0b013e3180316c92]
A., Ensini; Catani, Fabio; A., Leardini; M., Romagnoli; S., Giannini
File in questo prodotto:
File Dimensione Formato  
alignment and clinical results.pdf

Accesso riservato

Tipologia: Abstract
Dimensione 136.16 kB
Formato Adobe PDF
136.16 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/701557
Citazioni
  • ???jsp.display-item.citation.pmc??? 39
  • Scopus 172
  • ???jsp.display-item.citation.isi??? 146
social impact