PURPOSE: In some cases of total ankle replacement, perfect alignment of the prosthetic components is not achieved. This study analyses the extent to which component positioning is critical for the final range of motion. METHODS: Fourteen patients undergoing total ankle replacement were assessed preoperatively and postoperatively at seven and 13 months follow-up. X-ray pictures of the ankle were taken in static double leg stance, i.e. at neutral joint position, and in maximum plantarflexion and dorsiflexion. Measurements were obtained by a specially devised computer program based on anatomical reference points digitised on the radiograms. These allowed calculation of the position and orientation of the components in the sagittal and coronal planes, together with the joint range of motion. RESULTS: The mean range of motion was about 34 degrees at the first follow-up and maintained at the second. Tibial and talar components were more anterior than the mid-tibial shaft in 11 and nine patients, respectively. Mean inclination was about four degrees posterior for the tibial component and nearly one degree anterior for the talar component. A significantly larger range of motion was found in ankles both with the talar component located and inclined more anteriorly than the tibial. CONCLUSIONS: Correlation, though weak, was found between motion at the replaced ankle and possible residual subluxation and inclination of the components. However, a satisfactory range of motion was also achieved in those patients where recommended locations for the components could not be reached because of the size of the original joint deformity.

Position of the prosthesis components in total ankle replacement and the effect on motion at the replaced joint / F., Cenni; A., Leardini; A., Cheli; Catani, Fabio; C., Belvedere; M., Romagnoli; S., Giannini. - In: INTERNATIONAL ORTHOPAEDICS. - ISSN 0341-2695. - ELETTRONICO. - 7-2011:(2011), pp. 1-5. [10.1007/s00264-011-1323-6]

Position of the prosthesis components in total ankle replacement and the effect on motion at the replaced joint.

CATANI, Fabio;
2011

Abstract

PURPOSE: In some cases of total ankle replacement, perfect alignment of the prosthetic components is not achieved. This study analyses the extent to which component positioning is critical for the final range of motion. METHODS: Fourteen patients undergoing total ankle replacement were assessed preoperatively and postoperatively at seven and 13 months follow-up. X-ray pictures of the ankle were taken in static double leg stance, i.e. at neutral joint position, and in maximum plantarflexion and dorsiflexion. Measurements were obtained by a specially devised computer program based on anatomical reference points digitised on the radiograms. These allowed calculation of the position and orientation of the components in the sagittal and coronal planes, together with the joint range of motion. RESULTS: The mean range of motion was about 34 degrees at the first follow-up and maintained at the second. Tibial and talar components were more anterior than the mid-tibial shaft in 11 and nine patients, respectively. Mean inclination was about four degrees posterior for the tibial component and nearly one degree anterior for the talar component. A significantly larger range of motion was found in ankles both with the talar component located and inclined more anteriorly than the tibial. CONCLUSIONS: Correlation, though weak, was found between motion at the replaced ankle and possible residual subluxation and inclination of the components. However, a satisfactory range of motion was also achieved in those patients where recommended locations for the components could not be reached because of the size of the original joint deformity.
2011
7-2011
1
5
Position of the prosthesis components in total ankle replacement and the effect on motion at the replaced joint / F., Cenni; A., Leardini; A., Cheli; Catani, Fabio; C., Belvedere; M., Romagnoli; S., Giannini. - In: INTERNATIONAL ORTHOPAEDICS. - ISSN 0341-2695. - ELETTRONICO. - 7-2011:(2011), pp. 1-5. [10.1007/s00264-011-1323-6]
F., Cenni; A., Leardini; A., Cheli; Catani, Fabio; C., Belvedere; M., Romagnoli; S., Giannini
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/701541
Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 30
  • ???jsp.display-item.citation.isi??? 27
social impact