This study aimed to evaluate the accuracy of routine systems (Vitek2 cards AST-N022 and AST-N026; Kirby Bauer; Etest) for susceptibility testing of Pseudomonas aeruginosa to piperacillin and piperacillin-tazobactam. Vitek2 (card AST-N022) showed the worst performance; the other three methods (Vitek2 card AST-N026, Kirby-Bauer and E-test) performed comparably but never fulfilled the minimal standard proposed by FDA.
Accuracy of automated and manual systems for susceptibility testing of Pseudomonas aeruginosa to piperacillin and piperacillin-tazobactam / C., Gagliotti; M., Sarti; Sabia, Carla; R., Gargiulo; G. M., Rossolini; C., Carillo; C., Cassani; A. P., Cipolloni; F., Pedna; M. R., Rossi; S. S., Incerti; G., Testa; C., Venturelli; M. L., Moro. - In: NEW MICROBIOLOGICA. - ISSN 1121-7138. - ELETTRONICO. - 34:1(2011), pp. 97-99.
Accuracy of automated and manual systems for susceptibility testing of Pseudomonas aeruginosa to piperacillin and piperacillin-tazobactam.
SABIA, Carla;
2011
Abstract
This study aimed to evaluate the accuracy of routine systems (Vitek2 cards AST-N022 and AST-N026; Kirby Bauer; Etest) for susceptibility testing of Pseudomonas aeruginosa to piperacillin and piperacillin-tazobactam. Vitek2 (card AST-N022) showed the worst performance; the other three methods (Vitek2 card AST-N026, Kirby-Bauer and E-test) performed comparably but never fulfilled the minimal standard proposed by FDA.File | Dimensione | Formato | |
---|---|---|---|
newmicr_2011.pdf
Solo gestori archivio
Tipologia:
Versione pubblicata dall'editore
Dimensione
123.85 kB
Formato
Adobe PDF
|
123.85 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris