The design and realization of a new low-cost instrument for real-time measurement of blood flow particularly developed for extracorporeal blood circulator systems are presented. A self-mixing interferometric technique combined with a dedicated analog signal processing system was adopted to perform an accurate flow measurement. The developed system exhibits an accuracy lower than 3%. The cost of the entire measuring system has been estimated to be about 50 euros.
Low-Cost Optical Flowmeter With Analog Front-End Electronics for Blood Extracorporeal Circulators / M., Norgia; A., Pesatori; Rovati, Luigi. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - STAMPA. - 59:5(2010), pp. 1233-1239. [10.1109/TIM.2009.2038015]
Low-Cost Optical Flowmeter With Analog Front-End Electronics for Blood Extracorporeal Circulators
ROVATI, Luigi
2010
Abstract
The design and realization of a new low-cost instrument for real-time measurement of blood flow particularly developed for extracorporeal blood circulator systems are presented. A self-mixing interferometric technique combined with a dedicated analog signal processing system was adopted to perform an accurate flow measurement. The developed system exhibits an accuracy lower than 3%. The cost of the entire measuring system has been estimated to be about 50 euros.File | Dimensione | Formato | |
---|---|---|---|
05373834.pdf
Accesso riservato
Tipologia:
AAM - Versione dell'autore revisionata e accettata per la pubblicazione
Dimensione
1.29 MB
Formato
Adobe PDF
|
1.29 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris