Methods for evaluating the interface between bone and an experimental dental implant have been analyzed. The material studied was a titanium implant coated with a mixture of alumina and titanium dioxide by means of a plasma-jet system with the dog as the animal model. The evaluations were made on two levels: (1) in vivo, by analyzing radiographs of the peri-implant zone with a video display computer (2) in vitro, after explantation, by analyzing vertical and horizontal sections of the sample with an optical and a scanning electron microscope, and by a video display computer analysis of the microradiographs of these sections.
The evaluation of the interface between bone and a bioceramic dental implant / Gatti, Antonietta; Zaffe, Davide; G. P., Poli; Galetti, Riccardo. - In: JOURNAL OF BIOMEDICAL MATERIALS RESEARCH. - ISSN 0021-9304. - STAMPA. - 21:(1987), pp. 1005-1011. [10.1002/jbm.820210806]
The evaluation of the interface between bone and a bioceramic dental implant.
GATTI, Antonietta;ZAFFE, Davide;GALETTI, Riccardo
1987
Abstract
Methods for evaluating the interface between bone and an experimental dental implant have been analyzed. The material studied was a titanium implant coated with a mixture of alumina and titanium dioxide by means of a plasma-jet system with the dog as the animal model. The evaluations were made on two levels: (1) in vivo, by analyzing radiographs of the peri-implant zone with a video display computer (2) in vitro, after explantation, by analyzing vertical and horizontal sections of the sample with an optical and a scanning electron microscope, and by a video display computer analysis of the microradiographs of these sections.Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris