Vitallium appliances and surrounding tissues were investigated to evaluate the release and accumulation of elements.Four microplates, sixteen remaining screws and surrounding tissues were removed from three patients presenting inflammation 4 to 6 years after surgery for bone graft or fracture fixation and were submitted to SEM and X-ray microprobe analysis. Histology was performed on paraffin or PMMA sections of tissues.A continuous release of elements from metallic appliances into soft tissues was observed. Cobalt, chromium, and nickel were detected in soft and boney tissues in close proximity to the appliance. Aluminium, as a component of screw coatings, accumulated in soft tissues, and a remarkable amount of aluminium was detected in the dense lamella of lamellar bone.The results suggest that coatings containing aluminium should be avoided and the time these appliances are allowed to remain in patients should be shortened. Further studies on element release and the fate of aluminium in bone are warranted. IDS Number: 961PB PMID: 16167115

Release of elements from retrieved maxillofacial plates and screws / Bertoldi, Carlo; J. M., Pradelli; Consolo, Ugo; Zaffe, Davide. - In: JOURNAL OF MATERIALS SCIENCE. MATERIALS IN MEDICINE. - ISSN 0957-4530. - STAMPA. - 16:9(2005), pp. 857-861. [10.1007/s10856-005-3575-4]

Release of elements from retrieved maxillofacial plates and screws.

BERTOLDI, Carlo;CONSOLO, Ugo;ZAFFE, Davide
2005

Abstract

Vitallium appliances and surrounding tissues were investigated to evaluate the release and accumulation of elements.Four microplates, sixteen remaining screws and surrounding tissues were removed from three patients presenting inflammation 4 to 6 years after surgery for bone graft or fracture fixation and were submitted to SEM and X-ray microprobe analysis. Histology was performed on paraffin or PMMA sections of tissues.A continuous release of elements from metallic appliances into soft tissues was observed. Cobalt, chromium, and nickel were detected in soft and boney tissues in close proximity to the appliance. Aluminium, as a component of screw coatings, accumulated in soft tissues, and a remarkable amount of aluminium was detected in the dense lamella of lamellar bone.The results suggest that coatings containing aluminium should be avoided and the time these appliances are allowed to remain in patients should be shortened. Further studies on element release and the fate of aluminium in bone are warranted. IDS Number: 961PB PMID: 16167115
2005
16
9
857
861
Release of elements from retrieved maxillofacial plates and screws / Bertoldi, Carlo; J. M., Pradelli; Consolo, Ugo; Zaffe, Davide. - In: JOURNAL OF MATERIALS SCIENCE. MATERIALS IN MEDICINE. - ISSN 0957-4530. - STAMPA. - 16:9(2005), pp. 857-861. [10.1007/s10856-005-3575-4]
Bertoldi, Carlo; J. M., Pradelli; Consolo, Ugo; Zaffe, Davide
File in questo prodotto:
File Dimensione Formato  
JMaterMedic2005.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 803.27 kB
Formato Adobe PDF
803.27 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/595542
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 18
social impact