A finite element analysis has been performed for a hip replacement incorporating an elastomeric surface layer. The effects of assuming the elastomer as incompressible have been examined for imposed loads, by comparing the incompressible solution to that using accurately measured values for Poisson's ratio, and while deflections were found to be significantly different, the changes in maximum pressure in the contact, maximum shear stress, contact angle and angular position to maximum shear stress were all small (less than 10 per cent). Comparisons have also been made between the finite element results for ABAQUS and an asymptotic solution, and it was found to be important to use the numerical model particularly at small values of clearance

Axisymmetric Finite Element Analysis of a Hip Replacement with an Elastomeric Layer: the Effects of Clearance and Poisson's Ratio / A., Unsworth; Strozzi, Antonio. - In: PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE. - ISSN 0954-4119. - ELETTRONICO. - 209:(1995), pp. 59-64.

Axisymmetric Finite Element Analysis of a Hip Replacement with an Elastomeric Layer: the Effects of Clearance and Poisson's Ratio

STROZZI, Antonio
1995

Abstract

A finite element analysis has been performed for a hip replacement incorporating an elastomeric surface layer. The effects of assuming the elastomer as incompressible have been examined for imposed loads, by comparing the incompressible solution to that using accurately measured values for Poisson's ratio, and while deflections were found to be significantly different, the changes in maximum pressure in the contact, maximum shear stress, contact angle and angular position to maximum shear stress were all small (less than 10 per cent). Comparisons have also been made between the finite element results for ABAQUS and an asymptotic solution, and it was found to be important to use the numerical model particularly at small values of clearance
1995
209
59
64
Axisymmetric Finite Element Analysis of a Hip Replacement with an Elastomeric Layer: the Effects of Clearance and Poisson's Ratio / A., Unsworth; Strozzi, Antonio. - In: PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS. PART H, JOURNAL OF ENGINEERING IN MEDICINE. - ISSN 0954-4119. - ELETTRONICO. - 209:(1995), pp. 59-64.
A., Unsworth; Strozzi, Antonio
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/459239
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 12
  • ???jsp.display-item.citation.isi??? ND
social impact