Sirolimus is a immunosuppressive agent for renal transplant recipients. Monitoring of whole blood sirolimus concentration is necessary in order to improve clinical outcomes. An increasing number of clinical laboratories (4-14% during 2005) are using microparticle enzyme immunoassay (MEIA) for sirolimus quantitation but previous reports indicated a high variability, with a mean difference of 17% for MEIA method vs. high-performance liquid chromatography/ultraviolet (HPLC/UV). This study was aimed at comparing the reliability of MEIA with the HPLC/UV method. Blood samples from transplant patients were processed using both HPLC/UV and MEIA assays. Comparison and Bland-Altman plots, as well as regression analysis and paired t-test were used to compare results of the assays. Concentrations were stratified into three groups and used to investigate whether any observed difference between methods could be influenced by sirolimus concentration. Regression analysis yielded a coefficient of correlation R of 0.9756, the line of best fit being y = 0.9832x + 0.1976. The statistical analysis showed no difference between the two sets of experimental data. The average percentage difference between the two methods was found to be -0.2 +/- 19.2%. On the basis of our present results, the tested MEIA assay is able to quantify sirolimus concentration with a clinically acceptable imprecision, similar to that of HPLC/UV method.

Switching from HPLC/UV to MEIA for whole blood sirolimus quantitation: Comparison of methods / Pini, Luigi Alberto; Gallesi, Daniela; Brovia, Daria; Bertolini, Alfio; Pinetti, Diego; Ruggieri, Valentina; Pisa, S; Poppi, B; Castellana, Cn. - In: JOURNAL OF CLINICAL LABORATORY ANALYSIS. - ISSN 0887-8013. - STAMPA. - 20:6(2006), pp. 239-244. [10.1002/jcla.20148]

Switching from HPLC/UV to MEIA for whole blood sirolimus quantitation: Comparison of methods

PINI, Luigi Alberto;GALLESI, Daniela;BROVIA, Daria;BERTOLINI, Alfio;PINETTI, Diego;RUGGIERI, Valentina;
2006

Abstract

Sirolimus is a immunosuppressive agent for renal transplant recipients. Monitoring of whole blood sirolimus concentration is necessary in order to improve clinical outcomes. An increasing number of clinical laboratories (4-14% during 2005) are using microparticle enzyme immunoassay (MEIA) for sirolimus quantitation but previous reports indicated a high variability, with a mean difference of 17% for MEIA method vs. high-performance liquid chromatography/ultraviolet (HPLC/UV). This study was aimed at comparing the reliability of MEIA with the HPLC/UV method. Blood samples from transplant patients were processed using both HPLC/UV and MEIA assays. Comparison and Bland-Altman plots, as well as regression analysis and paired t-test were used to compare results of the assays. Concentrations were stratified into three groups and used to investigate whether any observed difference between methods could be influenced by sirolimus concentration. Regression analysis yielded a coefficient of correlation R of 0.9756, the line of best fit being y = 0.9832x + 0.1976. The statistical analysis showed no difference between the two sets of experimental data. The average percentage difference between the two methods was found to be -0.2 +/- 19.2%. On the basis of our present results, the tested MEIA assay is able to quantify sirolimus concentration with a clinically acceptable imprecision, similar to that of HPLC/UV method.
2006
20
6
239
244
Switching from HPLC/UV to MEIA for whole blood sirolimus quantitation: Comparison of methods / Pini, Luigi Alberto; Gallesi, Daniela; Brovia, Daria; Bertolini, Alfio; Pinetti, Diego; Ruggieri, Valentina; Pisa, S; Poppi, B; Castellana, Cn. - In: JOURNAL OF CLINICAL LABORATORY ANALYSIS. - ISSN 0887-8013. - STAMPA. - 20:6(2006), pp. 239-244. [10.1002/jcla.20148]
Pini, Luigi Alberto; Gallesi, Daniela; Brovia, Daria; Bertolini, Alfio; Pinetti, Diego; Ruggieri, Valentina; Pisa, S; Poppi, B; Castellana, Cn...espandi
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/22773
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 5
social impact