The in situ measurement of Sr isotopes in carbonates by MC-ICP-MS is limited by the availability of suitable microanalytical reference materials (RMs), which match the samples of interest. Whereas several well-characterised carbonate reference materials for Sr mass fractions > 1000 µg g−1 are available, there is a lack of well-characterised carbonate microanalytical RMs with lower Sr mass fractions. Here, we present a new synthetic carbonate nanopowder RM with a Sr mass fraction of ca. 500 µg g−1 suitable for microanalytical Sr isotope research (‘NanoSr’). NanoSr was analysed by both solution-based and in situ techniques. Element mass fractions were determined using EPMA (Ca mass fraction), as well as laser ablation and solution ICP-MS in different laboratories. The 87Sr/86Sr ratio was determined by well-established bulk methods for Sr isotope measurements and is 0.70756 ± 0.00003 (2s). The Sr isotope microhomogeneity of the material was determined by LA-MC-ICP-MS, which resulted in 87Sr/86Sr ratios of 0.70753 ± 0.00007 (2s) and 0.70757 ± 0.00006 (2s), respectively, in agreement with the solution data within uncertainties. Thus, this new reference material is well suited to monitor and correct microanalytical Sr isotope measurements of low-Sr, low-REE carbonate samples. NanoSr is available from the corresponding author.
NanoSr – A New Carbonate Microanalytical Reference Material for In Situ Strontium Isotope Analysis / Weber, M.; Lugli, F.; Hattendorf, B.; Scholz, D.; Mertz-Kraus, R.; Guinoiseau, D.; Jochum, K. P.. - In: GEOSTANDARDS AND GEOANALYTICAL RESEARCH. - ISSN 1639-4488. - 44:1(2020), pp. 69-83. [10.1111/ggr.12296]
NanoSr – A New Carbonate Microanalytical Reference Material for In Situ Strontium Isotope Analysis
Lugli F.;
2020
Abstract
The in situ measurement of Sr isotopes in carbonates by MC-ICP-MS is limited by the availability of suitable microanalytical reference materials (RMs), which match the samples of interest. Whereas several well-characterised carbonate reference materials for Sr mass fractions > 1000 µg g−1 are available, there is a lack of well-characterised carbonate microanalytical RMs with lower Sr mass fractions. Here, we present a new synthetic carbonate nanopowder RM with a Sr mass fraction of ca. 500 µg g−1 suitable for microanalytical Sr isotope research (‘NanoSr’). NanoSr was analysed by both solution-based and in situ techniques. Element mass fractions were determined using EPMA (Ca mass fraction), as well as laser ablation and solution ICP-MS in different laboratories. The 87Sr/86Sr ratio was determined by well-established bulk methods for Sr isotope measurements and is 0.70756 ± 0.00003 (2s). The Sr isotope microhomogeneity of the material was determined by LA-MC-ICP-MS, which resulted in 87Sr/86Sr ratios of 0.70753 ± 0.00007 (2s) and 0.70757 ± 0.00006 (2s), respectively, in agreement with the solution data within uncertainties. Thus, this new reference material is well suited to monitor and correct microanalytical Sr isotope measurements of low-Sr, low-REE carbonate samples. NanoSr is available from the corresponding author.File | Dimensione | Formato | |
---|---|---|---|
Geostandard Geoanalytic Res - 2019 - Weber - NanoSr A New Carbonate Microanalytical Reference Material for In Situ.pdf
Accesso riservato
Tipologia:
VOR - Versione pubblicata dall'editore
Dimensione
1.61 MB
Formato
Adobe PDF
|
1.61 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris