This study sought to develop a novel echocardiogram outflow ramp test to detect device malfunctions in centrifugal-flow left ventricular assist devices (LVADs). This new ramp pump test is based on the direct analyses of systolic and diastolic ratio (S/D) Doppler velocity in the outflow cannula in the HeartWare LVAD during progressive increases in speed. The results showed that in patients with normal pump function, the Doppler velocity S/D ratio gradually decreased during LVAD speed increases. This test is easily performed and seems promising to detect normal pump function in patients assisted by a centrifugal flow LVAD.
Echocardiographic outflow pump ramp test in centrifugal-flow left ventricular assist device / Iacovoni, A; Vittori, C; Fontana, A; Carobbio, A; Fino, C; D'Elia, E; Terzi, A; Senni, M. - In: THE INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS. - ISSN 1724-6040. - 40:3(2017), pp. 128-131. [10.5301/ijao.5000573]
Echocardiographic outflow pump ramp test in centrifugal-flow left ventricular assist device
Carobbio A;
2017
Abstract
This study sought to develop a novel echocardiogram outflow ramp test to detect device malfunctions in centrifugal-flow left ventricular assist devices (LVADs). This new ramp pump test is based on the direct analyses of systolic and diastolic ratio (S/D) Doppler velocity in the outflow cannula in the HeartWare LVAD during progressive increases in speed. The results showed that in patients with normal pump function, the Doppler velocity S/D ratio gradually decreased during LVAD speed increases. This test is easily performed and seems promising to detect normal pump function in patients assisted by a centrifugal flow LVAD.Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris