AIMS: The HeartLogic Index combines data from multiple implantable cardioverter defibrillators (ICDs) sensors and has been shown to accurately stratify patients at risk of heart failure (HF) events. We evaluated and compared the performance of this algorithm during sinus rhythm and during long-lasting atrial fibrillation (AF). METHODS AND RESULTS: HeartLogic was activated in 568 ICD patients from 26 centres. We found periods of ≥30 consecutive days with an atrial high-rate episode (AHRE) burden <1 h/day and periods with an AHRE burden ≥20 h/day. We then identified patients who met both criteria during the follow-up (AHRE group, n = 53), to allow pairwise comparison of periods. For control purposes, we identified patients with an AHRE burden <1 h throughout their follow-up and implemented 2:1 propensity score matching vs. the AHRE group (matched non-AHRE group, n = 106). In the AHRE group, the rate of alerts was 1.2 [95% confidence interval (CI): 1.0-1.5]/patient-year during periods with an AHRE burden <1 h/day and 2.0 (95% CI: 1.5-2.6)/patient-year during periods with an AHRE-burden ≥20 h/day (P = 0.004). The rate of HF hospitalizations was 0.34 (95% CI: 0.15-0.69)/patient-year during IN-alert periods and 0.06 (95% CI: 0.02-0.14)/patient-year during OUT-of-alert periods (P < 0.001). The IN/OUT-of-alert state incidence rate ratio of HF hospitalizations was 8.59 (95% CI: 1.67-55.31) during periods with an AHRE burden <1 h/day and 2.70 (95% CI: 1.01-28.33) during periods with an AHRE burden ≥20 h/day. In the matched non-AHRE group, the rate of HF hospitalizations was 0.29 (95% CI: 0.12-0.60)/patient-year during IN-alert periods and 0.04 (95% CI: 0.02-0.08)/patient-year during OUT-of-alert periods (P < 0.001). The incidence rate ratio was 7.11 (95% CI: 2.19-22.44). CONCLUSION: Patients received more alerts during periods of AF. The ability of the algorithm to identify increased risk of HF events was confirmed during AF, despite a lower IN/OUT-of-alert incidence rate ratio in comparison with non-AF periods and non-AF patients. CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/Identifier: NCT02275637.

Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation / Boriani, G.; Bertini, M.; Manzo, M.; Calo, L.; Santini, L.; Savarese, G.; Dello Russo, A.; Santobuono, V. E.; Lavalle, C.; Viscusi, M.; Amellone, C.; Calvanese, R.; Santoro, A.; Rapacciuolo, A.; Ziacchi, M.; Arena, G.; Imberti, J. F.; Campari, M.; Valsecchi, S.; D'Onofrio, A.. - In: EUROPACE. - ISSN 1532-2092. - 25:9(2023), pp. N/A-N/A. [10.1093/europace/euad261]

Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation

Boriani G.;Manzo M.;Imberti J. F.;
2023

Abstract

AIMS: The HeartLogic Index combines data from multiple implantable cardioverter defibrillators (ICDs) sensors and has been shown to accurately stratify patients at risk of heart failure (HF) events. We evaluated and compared the performance of this algorithm during sinus rhythm and during long-lasting atrial fibrillation (AF). METHODS AND RESULTS: HeartLogic was activated in 568 ICD patients from 26 centres. We found periods of ≥30 consecutive days with an atrial high-rate episode (AHRE) burden <1 h/day and periods with an AHRE burden ≥20 h/day. We then identified patients who met both criteria during the follow-up (AHRE group, n = 53), to allow pairwise comparison of periods. For control purposes, we identified patients with an AHRE burden <1 h throughout their follow-up and implemented 2:1 propensity score matching vs. the AHRE group (matched non-AHRE group, n = 106). In the AHRE group, the rate of alerts was 1.2 [95% confidence interval (CI): 1.0-1.5]/patient-year during periods with an AHRE burden <1 h/day and 2.0 (95% CI: 1.5-2.6)/patient-year during periods with an AHRE-burden ≥20 h/day (P = 0.004). The rate of HF hospitalizations was 0.34 (95% CI: 0.15-0.69)/patient-year during IN-alert periods and 0.06 (95% CI: 0.02-0.14)/patient-year during OUT-of-alert periods (P < 0.001). The IN/OUT-of-alert state incidence rate ratio of HF hospitalizations was 8.59 (95% CI: 1.67-55.31) during periods with an AHRE burden <1 h/day and 2.70 (95% CI: 1.01-28.33) during periods with an AHRE burden ≥20 h/day. In the matched non-AHRE group, the rate of HF hospitalizations was 0.29 (95% CI: 0.12-0.60)/patient-year during IN-alert periods and 0.04 (95% CI: 0.02-0.08)/patient-year during OUT-of-alert periods (P < 0.001). The incidence rate ratio was 7.11 (95% CI: 2.19-22.44). CONCLUSION: Patients received more alerts during periods of AF. The ability of the algorithm to identify increased risk of HF events was confirmed during AF, despite a lower IN/OUT-of-alert incidence rate ratio in comparison with non-AF periods and non-AF patients. CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/Identifier: NCT02275637.
2023
25
9
N/A
N/A
Performance of a multi-sensor implantable defibrillator algorithm for heart failure monitoring in the presence of atrial fibrillation / Boriani, G.; Bertini, M.; Manzo, M.; Calo, L.; Santini, L.; Savarese, G.; Dello Russo, A.; Santobuono, V. E.; Lavalle, C.; Viscusi, M.; Amellone, C.; Calvanese, R.; Santoro, A.; Rapacciuolo, A.; Ziacchi, M.; Arena, G.; Imberti, J. F.; Campari, M.; Valsecchi, S.; D'Onofrio, A.. - In: EUROPACE. - ISSN 1532-2092. - 25:9(2023), pp. N/A-N/A. [10.1093/europace/euad261]
Boriani, G.; Bertini, M.; Manzo, M.; Calo, L.; Santini, L.; Savarese, G.; Dello Russo, A.; Santobuono, V. E.; Lavalle, C.; Viscusi, M.; Amellone, C.; ...espandi
File in questo prodotto:
File Dimensione Formato  
euad261.pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 2.47 MB
Formato Adobe PDF
2.47 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1322379
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 2
social impact