Purpose: The Histolog® Scanner (SamanTree Medical SA, Lausanne, Switzerland) is a large field-of-view confocal laser scanning microscope designed to allow intraoperative margin assessment by the production of histological images ready for assessment in the operating room. We evaluated the feasibility and the performance of the Histolog® Scanner (HS) to correctly identify infiltrated margins in clinical practice of lumpectomy specimens. It was extrapolated if the utilization of the HS has the potential to reduce infiltrated margins and therefore reduce re-operation rates in patients undergoing breast conserving surgery (BCS) due to a primarily diagnosed breast cancer including ductal carcinoma in situ. Methods: This is a single-center, prospective, non-interventional, diagnostic pilot study including 50 consecutive patients receiving BCS. The complete surface of the specimen was scanned using the HS intraoperatively. The surgery and the intraoperative margin assessment of the specimen was performed according to the clinical routine consisting of conventional specimen radiography as well as the clinical impression of the surgeon. Three surgeons and an experienced pathologist assessed the scans produced by the HS for cancer cells on the surface. The potential of the HS to correctly identify involved margins was compared to the results of the conventional specimen radiography alone as well as the clinical routine. The histopathological report served as the gold standard. Results: 50 specimens corresponding to 300 surfaces were scanned by the HS. The mean sensitivity of the surgeons to identify involved margins with the HS was 37.5% ± 5.6%, the specificity was 75.2% ± 13.0%. The assessment of resection margins by the pathologist resulted in a sensitivity of 37.5% and a specificity of 81.0%, while the local clinical routine resulted in a sensitivity of 37.5% and a specificity of 78.2%. Conclusion: Acquisition of high-resolution histological images using the HS was feasible in clinical practice. Sensitivity and specificity were comparable to clinical routine. With more specific training and experience on image interpretation and acquisition, the HS may have the potential to enable more accuracy in the margin assessment of BCS specimens.

Imaging of lumpectomy surface with large field-of-view confocal laser scanning microscopy 'Histolog® scanner' for breast margin assessment in comparison with conventional specimen radiography / Togawa, Riku; Hederer, Johanna; Ragazzi, Moira; Bruckner, Thomas; Fastner, Sarah; Gomez, Christina; Hennigs, André; Nees, Juliane; Pfob, André; Riedel, Fabian; Schäfgen, Benedikt; Stieber, Anne; Lux, Michael P; Heil, Jörg; Golatta, Michael. - In: BREAST. - ISSN 1532-3080. - 68:(2023), pp. 194-200. [10.1016/j.breast.2023.02.010]

Imaging of lumpectomy surface with large field-of-view confocal laser scanning microscopy 'Histolog® scanner' for breast margin assessment in comparison with conventional specimen radiography

Ragazzi, Moira;
2023

Abstract

Purpose: The Histolog® Scanner (SamanTree Medical SA, Lausanne, Switzerland) is a large field-of-view confocal laser scanning microscope designed to allow intraoperative margin assessment by the production of histological images ready for assessment in the operating room. We evaluated the feasibility and the performance of the Histolog® Scanner (HS) to correctly identify infiltrated margins in clinical practice of lumpectomy specimens. It was extrapolated if the utilization of the HS has the potential to reduce infiltrated margins and therefore reduce re-operation rates in patients undergoing breast conserving surgery (BCS) due to a primarily diagnosed breast cancer including ductal carcinoma in situ. Methods: This is a single-center, prospective, non-interventional, diagnostic pilot study including 50 consecutive patients receiving BCS. The complete surface of the specimen was scanned using the HS intraoperatively. The surgery and the intraoperative margin assessment of the specimen was performed according to the clinical routine consisting of conventional specimen radiography as well as the clinical impression of the surgeon. Three surgeons and an experienced pathologist assessed the scans produced by the HS for cancer cells on the surface. The potential of the HS to correctly identify involved margins was compared to the results of the conventional specimen radiography alone as well as the clinical routine. The histopathological report served as the gold standard. Results: 50 specimens corresponding to 300 surfaces were scanned by the HS. The mean sensitivity of the surgeons to identify involved margins with the HS was 37.5% ± 5.6%, the specificity was 75.2% ± 13.0%. The assessment of resection margins by the pathologist resulted in a sensitivity of 37.5% and a specificity of 81.0%, while the local clinical routine resulted in a sensitivity of 37.5% and a specificity of 78.2%. Conclusion: Acquisition of high-resolution histological images using the HS was feasible in clinical practice. Sensitivity and specificity were comparable to clinical routine. With more specific training and experience on image interpretation and acquisition, the HS may have the potential to enable more accuracy in the margin assessment of BCS specimens.
2023
68
194
200
Imaging of lumpectomy surface with large field-of-view confocal laser scanning microscopy 'Histolog® scanner' for breast margin assessment in comparison with conventional specimen radiography / Togawa, Riku; Hederer, Johanna; Ragazzi, Moira; Bruckner, Thomas; Fastner, Sarah; Gomez, Christina; Hennigs, André; Nees, Juliane; Pfob, André; Riedel, Fabian; Schäfgen, Benedikt; Stieber, Anne; Lux, Michael P; Heil, Jörg; Golatta, Michael. - In: BREAST. - ISSN 1532-3080. - 68:(2023), pp. 194-200. [10.1016/j.breast.2023.02.010]
Togawa, Riku; Hederer, Johanna; Ragazzi, Moira; Bruckner, Thomas; Fastner, Sarah; Gomez, Christina; Hennigs, André; Nees, Juliane; Pfob, André; Riedel, Fabian; Schäfgen, Benedikt; Stieber, Anne; Lux, Michael P; Heil, Jörg; Golatta, Michael
File in questo prodotto:
File Dimensione Formato  
Histolog scanner_The Breast 68(2023).pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 4.72 MB
Formato Adobe PDF
4.72 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1302567
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 3
social impact