Background: The number of patients with cardiac implantable electronic devices (CIEDs) undergoing radiotherapy (RT) for cancer treatment is growing. At present, prevalence and predictors of RT-induced CIEDs malfunctions are not defined. Methods: Systematic review and meta-analysis conducted following the PRISMA recommendations. PubMed, Scopus and Google Scholar were searched from inception to 31/01/2022 for studies reporting RT-induced malfunctions in CIEDs patients. Aim was to assess the prevalence of RT-induced CIEDs malfunctions and identify potential predictors. Results: Thirty-two out of 3962 records matched the inclusion criteria and were included in the meta-analysis. A total of 135 CIEDs malfunctions were detected among 3121 patients (6.6%, 95% confidence interval [CI]: 5.1%–8.4%). The pooled prevalence increased moving from pacemaker (PM) to implantable cardioverter defibrillator (ICD), and cardiac resynchronization therapy and defibrillator (CRT-D) groups (4.1%, 95% CI: 2.9–5.8; 8.2% 95% CI: 5.9–11.3; and 19.8%, 95% CI: 11.4–32.2 respectively). A higher risk ratio (RR) of malfunctions was found when neutron-producing energies were used as compared to non-neutron-producing energies (RR 9.98, 95% CI: 5.09–19.60) and in patients with ICD/CRT-D as compared to patients with PM/CRT-P (RR 2.07, 95% CI: 1.40–3.06). On the contrary, no association was found between maximal radiation dose at CIED >2 Gy and CIEDs malfunctions (RR 0.93; 95% CI: 0.31–2.76). Conclusions: Radiotherapy related CIEDs malfunction had a prevalence ranging from 4% to 20%. The use of neutron-producing energies and more complex devices (ICD/CRT-D) were associated with higher risk of device malfunction, while the radiation dose at CIED did not significantly impact on the risk unless higher doses (>10 Gy) were used.
A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients / Malavasi, V. L.; Imberti, J. F.; Tosetti, A.; Romiti, G.; Vitolo, M.; Zecchin, M.; Mazzeo, E.; Giuseppina, D. M.; Lohr, F.; Lopez-Fernandez, T.; Boriani, G.. - In: EUROPEAN JOURNAL OF CLINICAL INVESTIGATION. - ISSN 0014-2972. - 53:1(2023), pp. N/A-N/A. [10.1111/eci.13862]
A systematic review and meta-analysis on oncological radiotherapy in patients with a cardiac implantable electronic device: Prevalence and predictors of device malfunction in 3121 patients
Vitolo M.;Lohr F.;Boriani G.
2023
Abstract
Background: The number of patients with cardiac implantable electronic devices (CIEDs) undergoing radiotherapy (RT) for cancer treatment is growing. At present, prevalence and predictors of RT-induced CIEDs malfunctions are not defined. Methods: Systematic review and meta-analysis conducted following the PRISMA recommendations. PubMed, Scopus and Google Scholar were searched from inception to 31/01/2022 for studies reporting RT-induced malfunctions in CIEDs patients. Aim was to assess the prevalence of RT-induced CIEDs malfunctions and identify potential predictors. Results: Thirty-two out of 3962 records matched the inclusion criteria and were included in the meta-analysis. A total of 135 CIEDs malfunctions were detected among 3121 patients (6.6%, 95% confidence interval [CI]: 5.1%–8.4%). The pooled prevalence increased moving from pacemaker (PM) to implantable cardioverter defibrillator (ICD), and cardiac resynchronization therapy and defibrillator (CRT-D) groups (4.1%, 95% CI: 2.9–5.8; 8.2% 95% CI: 5.9–11.3; and 19.8%, 95% CI: 11.4–32.2 respectively). A higher risk ratio (RR) of malfunctions was found when neutron-producing energies were used as compared to non-neutron-producing energies (RR 9.98, 95% CI: 5.09–19.60) and in patients with ICD/CRT-D as compared to patients with PM/CRT-P (RR 2.07, 95% CI: 1.40–3.06). On the contrary, no association was found between maximal radiation dose at CIED >2 Gy and CIEDs malfunctions (RR 0.93; 95% CI: 0.31–2.76). Conclusions: Radiotherapy related CIEDs malfunction had a prevalence ranging from 4% to 20%. The use of neutron-producing energies and more complex devices (ICD/CRT-D) were associated with higher risk of device malfunction, while the radiation dose at CIED did not significantly impact on the risk unless higher doses (>10 Gy) were used.File | Dimensione | Formato | |
---|---|---|---|
Eur J Clin Investigation - 2022 - Malavasi - A systematic review and meta‐analysis on oncological radiotherapy in patients.pdf
Open access
Tipologia:
Versione pubblicata dall'editore
Dimensione
1.98 MB
Formato
Adobe PDF
|
1.98 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris