Introduction. Ventriculoperitoneal shunt carries a non-negligible failure rate often requiring multiple surgical procedures during patient-s lifetime. The most common cause is obstruction of the ventricular catheter that can be embedded in adhesions with choroid plexus, ependymal tissue, and fibrous material. In such cases, or when an intraventricular tumor is present, particular attention must be paid when removal of the ventricular catheter is required. The potential adhesions with the tip of the ventricular catheter results in an increased risk of life-threatening hemorrhage. Materials and methods. We present the case of a ventriculoperitoneal shunt proximal revision in a von Hippel-Lindau affected patient. The neuroendoscopic exploration to restore the patency of the ventricular system made it possible to notice a fibrous adhesion between the tip of the catheter and a pituitary hemangioblastoma abutting into the third ventricle. Discussion. Pituitary stalk hemangioblastoma is an infrequent localization, although it represents the most common supratentorial site of hemangioblastoma in patients affected by von Hippel-Lindau syndrome. In this and in similar cases, endoscopic-assisted shunt revision allows visualizing the tip of the ventricular catheter and eventual adhesions that can be coagulated and cut by simple maneuvers, without any morbidity for patients. Conclusion. In a ventriculoperitoneal shunt revision with the concomitant presence of an intraventricular tumor or lesion at risk of bleeding, the surgical procedure should be performed under neuroendoscopic observation to reduce the risk of intraventricular hemorrhage.
Endoscopic-assisted removal of ventricular catheter in high-risk patients / Puzzolante, A.; Pavesi, G.; Feletti, A.. - In: NEJROHIRURGIÂ. - ISSN 2587-7569. - 23:4(2021), pp. 82-86. [10.17650/1683-3295-2021-23-4-82-86]
Endoscopic-assisted removal of ventricular catheter in high-risk patients
Puzzolante A.;Pavesi G.;
2021
Abstract
Introduction. Ventriculoperitoneal shunt carries a non-negligible failure rate often requiring multiple surgical procedures during patient-s lifetime. The most common cause is obstruction of the ventricular catheter that can be embedded in adhesions with choroid plexus, ependymal tissue, and fibrous material. In such cases, or when an intraventricular tumor is present, particular attention must be paid when removal of the ventricular catheter is required. The potential adhesions with the tip of the ventricular catheter results in an increased risk of life-threatening hemorrhage. Materials and methods. We present the case of a ventriculoperitoneal shunt proximal revision in a von Hippel-Lindau affected patient. The neuroendoscopic exploration to restore the patency of the ventricular system made it possible to notice a fibrous adhesion between the tip of the catheter and a pituitary hemangioblastoma abutting into the third ventricle. Discussion. Pituitary stalk hemangioblastoma is an infrequent localization, although it represents the most common supratentorial site of hemangioblastoma in patients affected by von Hippel-Lindau syndrome. In this and in similar cases, endoscopic-assisted shunt revision allows visualizing the tip of the ventricular catheter and eventual adhesions that can be coagulated and cut by simple maneuvers, without any morbidity for patients. Conclusion. In a ventriculoperitoneal shunt revision with the concomitant presence of an intraventricular tumor or lesion at risk of bleeding, the surgical procedure should be performed under neuroendoscopic observation to reduce the risk of intraventricular hemorrhage.File | Dimensione | Formato | |
---|---|---|---|
russian.pdf
Open access
Tipologia:
Versione pubblicata dall'editore
Dimensione
335.44 kB
Formato
Adobe PDF
|
335.44 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris