Background and aims Whether Fibroscan thresholds can be immediately adopted for none, some or all other shear wave elastography techniques has not been tested. The aim of the present study was to test the concordance of the findings obtained from 7 of the most recent ultrasound elastography machines with respect to Fibroscan. Methods Sixteen hepatitis C virus-related patients with fibrosis â¥2 and having reliable results at Fibroscan were investigated in two intercostal spaces using 7 different elastography machines. Coefficients of both precision (an index of data dispersion) and accuracy (an index of bias correction factors expressing different magnitudes of changes in comparison to the reference) were calculated. Results Median stiffness values differed among the different machines as did coefficients of both precision (range 0.54â0.72) and accuracy (range 0.28â0.87). When the average of the measurements of two intercostal spaces was considered, coefficients of precision significantly increased with all machines (range 0.72â0.90) whereas of accuracy improved more scatteredly and by a smaller degree (range 0.40â0.99). Conclusions The present results showed only moderate concordance of the majority of elastography machines with the Fibroscan results, preventing the possibility of the immediate universal adoption of Fibroscan thresholds for defining liver fibrosis staging for all new machines.
Differences in liver stiffness values obtained with new ultrasound elastography machines and Fibroscan: A comparative study / Piscaglia, Fabio; Salvatore, Veronica; Mulazzani, Lorenzo; Cantisani, Vito; Colecchia, Antonio; Di Donato, Roberto; Felicani, Cristina; Ferrarini, Alessia; Gamal, Nesrine; Grasso, Valentina; Marasco, Giovanni; Mazzotta, Elena; Ravaioli, Federico; Ruggieri, Giacomo; Serio, Ilaria; Sitouok Nkamgho, Joules Fabrice; Serra, Carla; Festi, Davide; Schiavone, Cosima; Bolondi, Luigi. - In: DIGESTIVE AND LIVER DISEASE. - ISSN 1590-8658. - 49:7(2017), pp. 802-808. [10.1016/j.dld.2017.03.001]
Differences in liver stiffness values obtained with new ultrasound elastography machines and Fibroscan: A comparative study
Colecchia, Antonio;Gamal, Nesrine;
2017
Abstract
Background and aims Whether Fibroscan thresholds can be immediately adopted for none, some or all other shear wave elastography techniques has not been tested. The aim of the present study was to test the concordance of the findings obtained from 7 of the most recent ultrasound elastography machines with respect to Fibroscan. Methods Sixteen hepatitis C virus-related patients with fibrosis â¥2 and having reliable results at Fibroscan were investigated in two intercostal spaces using 7 different elastography machines. Coefficients of both precision (an index of data dispersion) and accuracy (an index of bias correction factors expressing different magnitudes of changes in comparison to the reference) were calculated. Results Median stiffness values differed among the different machines as did coefficients of both precision (range 0.54â0.72) and accuracy (range 0.28â0.87). When the average of the measurements of two intercostal spaces was considered, coefficients of precision significantly increased with all machines (range 0.72â0.90) whereas of accuracy improved more scatteredly and by a smaller degree (range 0.40â0.99). Conclusions The present results showed only moderate concordance of the majority of elastography machines with the Fibroscan results, preventing the possibility of the immediate universal adoption of Fibroscan thresholds for defining liver fibrosis staging for all new machines.File | Dimensione | Formato | |
---|---|---|---|
Multielasto DLD.pdf
Accesso riservato
Dimensione
1.36 MB
Formato
Adobe PDF
|
1.36 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris