Aim: Ultrasound-based repositioning and real-time-monitoring aim at the improvement of the precision of SBRT in deep inspiration breath-hold (DIBH). Accuracy of ultrasound-based daily repositioning was estimated by comparison with DIBH-cone-beam-CT. Intrafraction motion during beam-delivery was assessed by ultrasound-real-time-monitoring. Patients/methods: Residual error after ultrasound-based interfractional repositioning (85 fractions, 16 SBRT-series; 14 patients) was assessed by marker-based (7 series) or liver-contour-based (9 series) matching in DIBH-CBCT. During beam-delivery, the percentage of 3D misalignment vector below 2 mm, between 2 and 5 mm, 5–7 mm and over 7 mm was estimated. Percentage of relevant target-displacements was analyzed as a function of DIBH-duration. Results: Residual error after ultrasound-based positioning was 0.4 ± 3.3 mm in LR (left-right), 0.2 ± 4.3 mm in CC (cranio-caudal) and 1.0 ± 3.0 mm in AP (anterior-posterior) directions (vector magnitude 5.4 ± 3.3 mm, MV ± SD). Over 544 DIBHs, target displacement was 1.3 ± 0.5 mm, 0.7 ± 0.3 mm, 1.6 ± 0.6 mm for CC, LR and AP directions, respectively (3D-vector 2.5 ± 0.7 mm). 3D misalignment vector length was below 2 mm in 49.8%, between 2 and 7 mm in 46.3%, and over 7 mm in 3.9% of the beam-delivery-time. During the first 5 s of the DIBH, 3D-misalignment vector length was always below 10 mm. Percentage of target displacements over 10 mm was 0.2%, 0.5% and 0.8% for 10 s, 15 s and 20 s DIBH-duration. Conclusions: Ultrasound-based interfractional repositioning is an accurate method for daily localization of abdominal DIBH-SBRT targets. Residual motion is <7 mm in 96% of the beam-delivery-time. Deviations >10 mm occur rarely and can be avoided by gating the beam at a predefined threshold. Ideal DIBH-duration should not exceed 15 s.

Ultrasound-based repositioning and real-time monitoring for abdominal SBRT in DIBH / Boda-Heggemann, J.; Sihono, D. S. K.; Streb, L.; Mertens, L.; Vogel, L.; Stieler, F.; Wenz, F.; Giordano, F. A.; Kalisch, I.; Lohr, F.; Fleckenstein, J.. - In: PHYSICA MEDICA. - ISSN 1120-1797. - 65:(2019), pp. 46-52. [10.1016/j.ejmp.2019.08.003]

Ultrasound-based repositioning and real-time monitoring for abdominal SBRT in DIBH

Lohr F.;
2019

Abstract

Aim: Ultrasound-based repositioning and real-time-monitoring aim at the improvement of the precision of SBRT in deep inspiration breath-hold (DIBH). Accuracy of ultrasound-based daily repositioning was estimated by comparison with DIBH-cone-beam-CT. Intrafraction motion during beam-delivery was assessed by ultrasound-real-time-monitoring. Patients/methods: Residual error after ultrasound-based interfractional repositioning (85 fractions, 16 SBRT-series; 14 patients) was assessed by marker-based (7 series) or liver-contour-based (9 series) matching in DIBH-CBCT. During beam-delivery, the percentage of 3D misalignment vector below 2 mm, between 2 and 5 mm, 5–7 mm and over 7 mm was estimated. Percentage of relevant target-displacements was analyzed as a function of DIBH-duration. Results: Residual error after ultrasound-based positioning was 0.4 ± 3.3 mm in LR (left-right), 0.2 ± 4.3 mm in CC (cranio-caudal) and 1.0 ± 3.0 mm in AP (anterior-posterior) directions (vector magnitude 5.4 ± 3.3 mm, MV ± SD). Over 544 DIBHs, target displacement was 1.3 ± 0.5 mm, 0.7 ± 0.3 mm, 1.6 ± 0.6 mm for CC, LR and AP directions, respectively (3D-vector 2.5 ± 0.7 mm). 3D misalignment vector length was below 2 mm in 49.8%, between 2 and 7 mm in 46.3%, and over 7 mm in 3.9% of the beam-delivery-time. During the first 5 s of the DIBH, 3D-misalignment vector length was always below 10 mm. Percentage of target displacements over 10 mm was 0.2%, 0.5% and 0.8% for 10 s, 15 s and 20 s DIBH-duration. Conclusions: Ultrasound-based interfractional repositioning is an accurate method for daily localization of abdominal DIBH-SBRT targets. Residual motion is <7 mm in 96% of the beam-delivery-time. Deviations >10 mm occur rarely and can be avoided by gating the beam at a predefined threshold. Ideal DIBH-duration should not exceed 15 s.
2019
65
46
52
Ultrasound-based repositioning and real-time monitoring for abdominal SBRT in DIBH / Boda-Heggemann, J.; Sihono, D. S. K.; Streb, L.; Mertens, L.; Vogel, L.; Stieler, F.; Wenz, F.; Giordano, F. A.; Kalisch, I.; Lohr, F.; Fleckenstein, J.. - In: PHYSICA MEDICA. - ISSN 1120-1797. - 65:(2019), pp. 46-52. [10.1016/j.ejmp.2019.08.003]
Boda-Heggemann, J.; Sihono, D. S. K.; Streb, L.; Mertens, L.; Vogel, L.; Stieler, F.; Wenz, F.; Giordano, F. A.; Kalisch, I.; Lohr, F.; Fleckenstein, J.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1222240
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 8
social impact