We report on the development of a multilayer coated reflection-type long period fiber grating (LPG) biosensor, useful for the detection of antibiotic resistance bacteria. A standard LPG is first transformed in a more practical probe working in reflection mode, then it is coated by a primary layer of aPS and a secondary layer of PMMA in order to increase its surrounding refractive index sensitivity and at the same time provide the necessary conditions for a correct biofunctionalization. Standard linkage chemistry has been applied to anchor the bioreceptors on the probe surface. We show some preliminary results demonstrating the capability of our LPG biosensor to successfully monitor all the biological steps of the biomolecular experiments, including β-lactamase binding detection tests. © 2014 SPIE.

Reflection-type long period grating biosensor for the detection of drug resistant bacteria: The Opto-bacteria Project / Consales, M.; Quero, G.; Zuppolini, S.; Sansone, L.; Borriello, A.; Giordano, M.; Venturelli, A.; Cusano, A.. - 9157:(2014), p. 91575G. (Intervento presentato al convegno 23rd International Conference on Optical Fibre Sensors tenutosi a Santander, esp nel 2014) [10.1117/12.2059783].

Reflection-type long period grating biosensor for the detection of drug resistant bacteria: The Opto-bacteria Project

Venturelli A.;
2014

Abstract

We report on the development of a multilayer coated reflection-type long period fiber grating (LPG) biosensor, useful for the detection of antibiotic resistance bacteria. A standard LPG is first transformed in a more practical probe working in reflection mode, then it is coated by a primary layer of aPS and a secondary layer of PMMA in order to increase its surrounding refractive index sensitivity and at the same time provide the necessary conditions for a correct biofunctionalization. Standard linkage chemistry has been applied to anchor the bioreceptors on the probe surface. We show some preliminary results demonstrating the capability of our LPG biosensor to successfully monitor all the biological steps of the biomolecular experiments, including β-lactamase binding detection tests. © 2014 SPIE.
2014
23rd International Conference on Optical Fibre Sensors
Santander, esp
2014
9157
91575G
Consales, M.; Quero, G.; Zuppolini, S.; Sansone, L.; Borriello, A.; Giordano, M.; Venturelli, A.; Cusano, A.
Reflection-type long period grating biosensor for the detection of drug resistant bacteria: The Opto-bacteria Project / Consales, M.; Quero, G.; Zuppolini, S.; Sansone, L.; Borriello, A.; Giordano, M.; Venturelli, A.; Cusano, A.. - 9157:(2014), p. 91575G. (Intervento presentato al convegno 23rd International Conference on Optical Fibre Sensors tenutosi a Santander, esp nel 2014) [10.1117/12.2059783].
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris