Purpose: To report midterm outcomes of the Nellix Endovascular Aneurysm Sealing (EVAS) System in the treatment of abdominal aortic aneurysm (AAA). Methods: Between September 2013 and July 2014, 64 AAA patients (mean age 76.6±6.8 years; 61 men) were treated with the EVAS system at 2 centers (only procedures performed at least 12 months prior to the analysis were included). Most patients were treated for a stable AAA, while 1 patient was treated for a ruptured aneurysm. Mean aneurysm diameter was 57.3±9.3 mm. The proximal neck measured a mean 21.5±3.3 mm in diameter and 27.0±12.1 mm long; the neck angle was 16.9°±19.3°. Eleven (17.2%) patients were treated outside the instructions for use (IFU). Results: Technical success was achieved in 63 (98.4%) of 64 patients; 1 type Ia endoleak was treated intraoperatively. One (1.6%) aneurysm-related death occurred at 4 months due to a secondary aortoenteric fistula. Overall, endoleaks occurred in 3 (4.7%) patients (2 type Ia, 1 type II). The estimated rates for 18-month overall survival, freedom from aneurysm-related death, and freedom from secondary interventions were 92.7%, 98.4%, and 95.0%, respectively. Patients treated outside the IFU had a significantly higher incidence of device-related complications (p=0.03). Conclusion: The use of the Nellix device in everyday clinical practice is safe and offers promising midterm results. The risk of secondary aortoenteric fistula requires further analysis. Longer follow-up is needed to assess the actual efficacy of the device, although the risk of migration with late endoleak seems low.
Midterm Outcomes of the Nellix Endovascular Aneurysm Sealing System: A Dual-Center Experience / Silingardi, R.; Coppi, G.; Ferrero, E.; Lauricella, A.; Psacharopulo, D.; Saitta, G.; Viazzo, A.; Ferri, M.. - In: JOURNAL OF ENDOVASCULAR THERAPY. - ISSN 1526-6028. - 23:5(2016), pp. 695-700. [10.1177/1526602816656355]
Midterm Outcomes of the Nellix Endovascular Aneurysm Sealing System: A Dual-Center Experience
Silingardi R.;Coppi G.
;
2016
Abstract
Purpose: To report midterm outcomes of the Nellix Endovascular Aneurysm Sealing (EVAS) System in the treatment of abdominal aortic aneurysm (AAA). Methods: Between September 2013 and July 2014, 64 AAA patients (mean age 76.6±6.8 years; 61 men) were treated with the EVAS system at 2 centers (only procedures performed at least 12 months prior to the analysis were included). Most patients were treated for a stable AAA, while 1 patient was treated for a ruptured aneurysm. Mean aneurysm diameter was 57.3±9.3 mm. The proximal neck measured a mean 21.5±3.3 mm in diameter and 27.0±12.1 mm long; the neck angle was 16.9°±19.3°. Eleven (17.2%) patients were treated outside the instructions for use (IFU). Results: Technical success was achieved in 63 (98.4%) of 64 patients; 1 type Ia endoleak was treated intraoperatively. One (1.6%) aneurysm-related death occurred at 4 months due to a secondary aortoenteric fistula. Overall, endoleaks occurred in 3 (4.7%) patients (2 type Ia, 1 type II). The estimated rates for 18-month overall survival, freedom from aneurysm-related death, and freedom from secondary interventions were 92.7%, 98.4%, and 95.0%, respectively. Patients treated outside the IFU had a significantly higher incidence of device-related complications (p=0.03). Conclusion: The use of the Nellix device in everyday clinical practice is safe and offers promising midterm results. The risk of secondary aortoenteric fistula requires further analysis. Longer follow-up is needed to assess the actual efficacy of the device, although the risk of migration with late endoleak seems low.File | Dimensione | Formato | |
---|---|---|---|
silingardi2016.pdf
Accesso riservato
Tipologia:
VOR - Versione pubblicata dall'editore
Dimensione
387.4 kB
Formato
Adobe PDF
|
387.4 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris