INTRODUCTION: The aim of this study was to compare the influences from different access angles and curvature radii on cyclic fatigue resistance of nickel-titanium rotary files. METHODS: Two file systems (2Shape [TS; MicroMega, Besançon, France] and HyFlex CM [HCM; Coltène/Whaledent, Allstätten, Switzerland]) were used. A total of 192 instruments of TS #25/.04 (TS1), TS #25/.06 (TS2), HCM #25/.04, and HCM #25/.06 were evaluated at 3 insertion angles (0°, 10°, and 20°) and 2 radii (5 mm and 3 mm) in 16-mm stainless steel artificial canals with a 60° curvature. Cyclic fatigue resistance was determined by the number of cycles to failure (NCF) using a customized testing device. Data were analyzed statistically with the significance level established at 95%. RESULTS: In the 3-mm radius canal, the instruments showed lower cyclic fatigue resistance than in the 5-mm radius canal (P < .05). HCM #25/.06 and all .04 taper instruments had a significant NCF reduction at 20° and 10° in the 3-mm radius canal (P < .05), whereas TS2 showed no significant differences. In the 5-mm radius of curvature, although .06 taper instruments had no significant NCF reduction for each angle tested, .04 taper files exhibited significant NCF reduction when tested at 20° (P < .05). Comparing the same size instruments, HCM had higher NCF than TS (P < .05). Instruments with a .04 taper exhibited higher NCF than the .06 ones with the same heat treatment (P < .05). CONCLUSIONS: An inclined insertion into the canals decreased cyclic fatigue resistance of thermal-treated instruments with a .04 taper at all radii of curvature tested. The synergistic effect of a small radius of curvature and access angulation of heat-treated instruments decreases their fatigue resistance.

Cyclic fatigue resistance of nickel-titanium rotary instruments according to the angle of file access and radius of root canal / Pedullà, E; La Rosa, Grm; Virgilio, C; Rapisarda, E; Kim, Hc; Generali, L. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 46:3(2020), pp. 431-436. [10.1016/j.joen.2019.11.015]

Cyclic fatigue resistance of nickel-titanium rotary instruments according to the angle of file access and radius of root canal.

Generali L
2020

Abstract

INTRODUCTION: The aim of this study was to compare the influences from different access angles and curvature radii on cyclic fatigue resistance of nickel-titanium rotary files. METHODS: Two file systems (2Shape [TS; MicroMega, Besançon, France] and HyFlex CM [HCM; Coltène/Whaledent, Allstätten, Switzerland]) were used. A total of 192 instruments of TS #25/.04 (TS1), TS #25/.06 (TS2), HCM #25/.04, and HCM #25/.06 were evaluated at 3 insertion angles (0°, 10°, and 20°) and 2 radii (5 mm and 3 mm) in 16-mm stainless steel artificial canals with a 60° curvature. Cyclic fatigue resistance was determined by the number of cycles to failure (NCF) using a customized testing device. Data were analyzed statistically with the significance level established at 95%. RESULTS: In the 3-mm radius canal, the instruments showed lower cyclic fatigue resistance than in the 5-mm radius canal (P < .05). HCM #25/.06 and all .04 taper instruments had a significant NCF reduction at 20° and 10° in the 3-mm radius canal (P < .05), whereas TS2 showed no significant differences. In the 5-mm radius of curvature, although .06 taper instruments had no significant NCF reduction for each angle tested, .04 taper files exhibited significant NCF reduction when tested at 20° (P < .05). Comparing the same size instruments, HCM had higher NCF than TS (P < .05). Instruments with a .04 taper exhibited higher NCF than the .06 ones with the same heat treatment (P < .05). CONCLUSIONS: An inclined insertion into the canals decreased cyclic fatigue resistance of thermal-treated instruments with a .04 taper at all radii of curvature tested. The synergistic effect of a small radius of curvature and access angulation of heat-treated instruments decreases their fatigue resistance.
2020
3-gen-2020
46
3
431
436
Cyclic fatigue resistance of nickel-titanium rotary instruments according to the angle of file access and radius of root canal / Pedullà, E; La Rosa, Grm; Virgilio, C; Rapisarda, E; Kim, Hc; Generali, L. - In: JOURNAL OF ENDODONTICS. - ISSN 0099-2399. - 46:3(2020), pp. 431-436. [10.1016/j.joen.2019.11.015]
Pedullà, E; La Rosa, Grm; Virgilio, C; Rapisarda, E; Kim, Hc; Generali, L
File in questo prodotto:
File Dimensione Formato  
Pedullà et al. 20202.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 1.17 MB
Formato Adobe PDF
1.17 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1198336
Citazioni
  • ???jsp.display-item.citation.pmc??? 9
  • Scopus 41
  • ???jsp.display-item.citation.isi??? 31
social impact