The number of patients with cardiac implantable electronic devices (CIEDs) requiring radiation therapy (RT) for cancer treatment is increasing. The purpose of this study is to estimate the prevalence, possible predictors, and clinical impact of RT-related CIEDs malfunctions. We retrospectively reviewed the medical records of all pacemaker (PM)/implantable cardioverter-defibrillator (ICD) patients who underwent RT in the last 14 years. One hundred and twenty-seven patients who underwent 150 separate RT courses were analysed (99 with a PM and 27 with an ICD). Of note, 21/127 (16.6%) patients were PM-dependent. Neutron-producing RT was used in 37/139 (26.6%) courses, whereas non-neutron-producing RT was used in 102/139 (73.4%) courses. The cumulative dose (Dmax) delivered to the CIED exceeded 5 Gy only in 2/132 (1.5%) cases. Device malfunctions were observed in 3/150 (2%) RT courses, but none was life-threatening or led to a major clinical event and all were resolved by CIED reprogramming. In all cases, the Dmax delivered to the CIED was < 2 Gy. Two malfunctions occurred in the 37 patients treated with neutron-producing RT (5.4%), and 1 malfunction occurred in the 102 patients treated with non-neutron-producing RT (1%) (p = 0.17). Device relocation from the RT field was performed in 2/127 (1.6%) patients. RT in patients with CIED is substantially safe if performed in an appropriately organized environment, with uncommon CIEDs malfunctions and no major clinical events. Neutron-producing energies, rather than Dmax, seem to increase the risk of malfunctions. Device interrogation on a regular basis is advised to promptly manage CIED malfunctions.

Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients / Malavasi, V. L.; De Marco, G.; Imberti, J. F.; Placentino, F.; Vitolo, M.; Mazzeo, E.; Cicoria, G.; Casali, E.; Turco, V.; Lohr, F.; Boriani, G.. - In: INTERNAL AND EMERGENCY MEDICINE. - ISSN 1828-0447. - 15:6(2020), pp. 967-973. [10.1007/s11739-019-02240-y]

Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients

Imberti J. F.;Placentino F.;Vitolo M.;Lohr F.;Boriani G.
2020

Abstract

The number of patients with cardiac implantable electronic devices (CIEDs) requiring radiation therapy (RT) for cancer treatment is increasing. The purpose of this study is to estimate the prevalence, possible predictors, and clinical impact of RT-related CIEDs malfunctions. We retrospectively reviewed the medical records of all pacemaker (PM)/implantable cardioverter-defibrillator (ICD) patients who underwent RT in the last 14 years. One hundred and twenty-seven patients who underwent 150 separate RT courses were analysed (99 with a PM and 27 with an ICD). Of note, 21/127 (16.6%) patients were PM-dependent. Neutron-producing RT was used in 37/139 (26.6%) courses, whereas non-neutron-producing RT was used in 102/139 (73.4%) courses. The cumulative dose (Dmax) delivered to the CIED exceeded 5 Gy only in 2/132 (1.5%) cases. Device malfunctions were observed in 3/150 (2%) RT courses, but none was life-threatening or led to a major clinical event and all were resolved by CIED reprogramming. In all cases, the Dmax delivered to the CIED was < 2 Gy. Two malfunctions occurred in the 37 patients treated with neutron-producing RT (5.4%), and 1 malfunction occurred in the 102 patients treated with non-neutron-producing RT (1%) (p = 0.17). Device relocation from the RT field was performed in 2/127 (1.6%) patients. RT in patients with CIED is substantially safe if performed in an appropriately organized environment, with uncommon CIEDs malfunctions and no major clinical events. Neutron-producing energies, rather than Dmax, seem to increase the risk of malfunctions. Device interrogation on a regular basis is advised to promptly manage CIED malfunctions.
2020
2-dic-2019
15
6
967
973
Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients / Malavasi, V. L.; De Marco, G.; Imberti, J. F.; Placentino, F.; Vitolo, M.; Mazzeo, E.; Cicoria, G.; Casali, E.; Turco, V.; Lohr, F.; Boriani, G.. - In: INTERNAL AND EMERGENCY MEDICINE. - ISSN 1828-0447. - 15:6(2020), pp. 967-973. [10.1007/s11739-019-02240-y]
Malavasi, V. L.; De Marco, G.; Imberti, J. F.; Placentino, F.; Vitolo, M.; Mazzeo, E.; Cicoria, G.; Casali, E.; Turco, V.; Lohr, F.; Boriani, G....espandi
File in questo prodotto:
File Dimensione Formato  
Malavasi2020_Article_Radiotherapy-inducedMalfunctio.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 555.69 kB
Formato Adobe PDF
555.69 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1190409
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 13
social impact