In this paper, we present a novel hyperspectral system designed for imaging the human iris in vivo. The instrument is based on a tunable light source (from 480 nm to 900 nm) and a monochrome camera. The system can acquire hyperspectral cubes in 4 s, with spatial and spectral resolutions of 11 μm and 20 nm, respectively. The optical characterization of the new instrument is described, and we demonstrate its safety with respect to ANSI standards. Preliminary in vivo measurements in human volunteers are also presented.
Design and performance of a hyperspectral imaging system: Preliminary in vivo spectral reflectance measurements of the human iris / Di Cecilia, L.; Rovati, L.. - In: REVIEW OF SCIENTIFIC INSTRUMENTS. - ISSN 0034-6748. - 91:1(2020), pp. 014104-014104-11. [10.1063/1.5125575]
Design and performance of a hyperspectral imaging system: Preliminary in vivo spectral reflectance measurements of the human iris
Di Cecilia, L.;Rovati, L.
2020
Abstract
In this paper, we present a novel hyperspectral system designed for imaging the human iris in vivo. The instrument is based on a tunable light source (from 480 nm to 900 nm) and a monochrome camera. The system can acquire hyperspectral cubes in 4 s, with spatial and spectral resolutions of 11 μm and 20 nm, respectively. The optical characterization of the new instrument is described, and we demonstrate its safety with respect to ANSI standards. Preliminary in vivo measurements in human volunteers are also presented.File | Dimensione | Formato | |
---|---|---|---|
RSI2020.pdf
Accesso riservato
Tipologia:
Versione pubblicata dall'editore
Dimensione
5.94 MB
Formato
Adobe PDF
|
5.94 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris