Background: Interstitial lung disease (ILD) is a severe systemic manifestation of rheumatoid arthritis (RA). High-resolution computed tomography (HRCT) represents the gold standard for the diagnosis of ILD, but its routine use for screening programs is not advisable because of both high cost and X-ray exposure. Velcro crackles at lung auscultation occur very early in the course of interstitial pneumonia, and their detection is an indication for HRCT. Recently, we developed an algorithm (VECTOR) to detect the presence of Velcro crackles in pulmonary sounds and showed good results in a small sample of RA patients. The aim of the present investigation was to validate the diagnostic accuracy of VECTOR in a larger population of RA patients, compared with that of the reference standard of HRCT, from a multicentre study. Methods: To avoid X-ray exposure, we enrolled 137 consecutive RA patients who had recently undergone HRCT. Lung sounds of all patients were recorded in 4 pulmonary fields bilaterally with a commercial electronic stethoscope (ES); subsequently, all HRCT images were blindly evaluated by a radiologist, and audio data were analysed by means of VECTOR. Results: Fifty-nine of 137 patients showed ILD (43.1%). VECTOR correctly classified 115/137 patients, showing a diagnostic accuracy of 83.9% and a sensitivity and specificity of 93.2 and 76.9%, respectively. Conclusions: VECTOR may represent the first validated tool for the screening of RA patients who are suspected for ILD and who should be directed to HRCT for the diagnosis. Moreover, early identification of RA-ILD could contribute to the design of prospective studies aimed at elucidating unclear aspects of the disease.

Diagnostic accuracy of a velcro sound detector (VECTOR) for interstitial lung disease in rheumatoid arthritis patients: The InSPIRAtE validation study (INterStitial pneumonia in rheumatoid ArThritis with an electronic device) / Manfredi, A.; Cassone, G.; Cerri, S.; Venerito, V.; Fedele, A. L.; Trevisani, M.; Furini, F.; Addimanda, O.; Pancaldi, F.; Della Casa, G.; D'Amico, R.; Vicini, R.; Sandri, G.; Torricelli, P.; Celentano, I.; Bortoluzzi, A.; Malavolta, N.; Meliconi, R.; Iannone, F.; Gremese, E.; Luppi, F.; Salvarani, C.; Sebastiani, M.. - In: BMC PULMONARY MEDICINE. - ISSN 1471-2466. - 19:1(2019), pp. 111-117. [10.1186/s12890-019-0875-x]

Diagnostic accuracy of a velcro sound detector (VECTOR) for interstitial lung disease in rheumatoid arthritis patients: The InSPIRAtE validation study (INterStitial pneumonia in rheumatoid ArThritis with an electronic device)

Manfredi A.
;
Cassone G.;Cerri S.;Pancaldi F.;Della Casa G.;D'Amico R.;Vicini R.;Sandri G.;Torricelli P.;Celentano I.;Salvarani C.;Sebastiani M.
2019

Abstract

Background: Interstitial lung disease (ILD) is a severe systemic manifestation of rheumatoid arthritis (RA). High-resolution computed tomography (HRCT) represents the gold standard for the diagnosis of ILD, but its routine use for screening programs is not advisable because of both high cost and X-ray exposure. Velcro crackles at lung auscultation occur very early in the course of interstitial pneumonia, and their detection is an indication for HRCT. Recently, we developed an algorithm (VECTOR) to detect the presence of Velcro crackles in pulmonary sounds and showed good results in a small sample of RA patients. The aim of the present investigation was to validate the diagnostic accuracy of VECTOR in a larger population of RA patients, compared with that of the reference standard of HRCT, from a multicentre study. Methods: To avoid X-ray exposure, we enrolled 137 consecutive RA patients who had recently undergone HRCT. Lung sounds of all patients were recorded in 4 pulmonary fields bilaterally with a commercial electronic stethoscope (ES); subsequently, all HRCT images were blindly evaluated by a radiologist, and audio data were analysed by means of VECTOR. Results: Fifty-nine of 137 patients showed ILD (43.1%). VECTOR correctly classified 115/137 patients, showing a diagnostic accuracy of 83.9% and a sensitivity and specificity of 93.2 and 76.9%, respectively. Conclusions: VECTOR may represent the first validated tool for the screening of RA patients who are suspected for ILD and who should be directed to HRCT for the diagnosis. Moreover, early identification of RA-ILD could contribute to the design of prospective studies aimed at elucidating unclear aspects of the disease.
2019
19
1
111
117
Diagnostic accuracy of a velcro sound detector (VECTOR) for interstitial lung disease in rheumatoid arthritis patients: The InSPIRAtE validation study (INterStitial pneumonia in rheumatoid ArThritis with an electronic device) / Manfredi, A.; Cassone, G.; Cerri, S.; Venerito, V.; Fedele, A. L.; Trevisani, M.; Furini, F.; Addimanda, O.; Pancaldi, F.; Della Casa, G.; D'Amico, R.; Vicini, R.; Sandri, G.; Torricelli, P.; Celentano, I.; Bortoluzzi, A.; Malavolta, N.; Meliconi, R.; Iannone, F.; Gremese, E.; Luppi, F.; Salvarani, C.; Sebastiani, M.. - In: BMC PULMONARY MEDICINE. - ISSN 1471-2466. - 19:1(2019), pp. 111-117. [10.1186/s12890-019-0875-x]
Manfredi, A.; Cassone, G.; Cerri, S.; Venerito, V.; Fedele, A. L.; Trevisani, M.; Furini, F.; Addimanda, O.; Pancaldi, F.; Della Casa, G.; D'Amico, R....espandi
File in questo prodotto:
File Dimensione Formato  
Manfredi-2019-Diagnostic-accuracy-of-a-velcro-sou.pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 640.49 kB
Formato Adobe PDF
640.49 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1179737
Citazioni
  • ???jsp.display-item.citation.pmc??? 26
  • Scopus 50
  • ???jsp.display-item.citation.isi??? 48
social impact