Among the numerous thermotherapy methods, heat wraps have been largely used over the last 2 decades as a self-administered practice for pain relief. Therefore, understanding their performance has become instrumental within the healthcare industry. However, the majority of the available studies have been focused on in vivo clinical performance, whereas a standardized, quantitative approach to evaluate and compare the various heat-wrap types against each other is lacking. An experimental methodology is proposed to carry out a comparative assessment between heat wraps in terms of their transient thermal behavior. A simple setup was developed to measure wrap/substrate interface temperature trend. The approach was validated by a preliminary infrared-thermography assessment and statistical analysis on the extensive dataset acquired on commercial heat-wrap types for low-back and neck pain relief. The heat-release trend was found to be qualitatively similar over all the investigated types, consisting of rapid growth, stationary phase, decay and end of the reactions. A set of parameters is also proposed to sum-marize heat-wrap thermal performance.
Experimental methodology for quantitative assessment of heat-wrap thermal transient behavior / Santangelo, Paolo Emilio; Santunione, Giulia; Muscio, Alberto. - In: MEDICAL ENGINEERING & PHYSICS. - ISSN 1873-4030. - 69:(2019), pp. 72-84. [10.1016/j.medengphy.2019.05.003]
Experimental methodology for quantitative assessment of heat-wrap thermal transient behavior
Paolo Emilio Santangelo
;Giulia Santunione;Alberto Muscio
2019
Abstract
Among the numerous thermotherapy methods, heat wraps have been largely used over the last 2 decades as a self-administered practice for pain relief. Therefore, understanding their performance has become instrumental within the healthcare industry. However, the majority of the available studies have been focused on in vivo clinical performance, whereas a standardized, quantitative approach to evaluate and compare the various heat-wrap types against each other is lacking. An experimental methodology is proposed to carry out a comparative assessment between heat wraps in terms of their transient thermal behavior. A simple setup was developed to measure wrap/substrate interface temperature trend. The approach was validated by a preliminary infrared-thermography assessment and statistical analysis on the extensive dataset acquired on commercial heat-wrap types for low-back and neck pain relief. The heat-release trend was found to be qualitatively similar over all the investigated types, consisting of rapid growth, stationary phase, decay and end of the reactions. A set of parameters is also proposed to sum-marize heat-wrap thermal performance.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S135045331930089X-main.pdf
Accesso riservato
Tipologia:
Versione pubblicata dall'editore
Dimensione
2.51 MB
Formato
Adobe PDF
|
2.51 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris