Background: Sleep apnea (SA) is a relevant issue in the management of patients with heart failure for risk stratification and for implementing treatment strategies. Objective: The purpose of this study was to evaluate in patients with implantable cardioverter-defibrillators (ICDs) the performance of the respiratory disturbance index (RDI) computed by the ApneaScan algorithm (Boston Scientific Inc., Natick, MA) as a discriminator of severe SA. Methods: ICD-indicated patients with left ventricular ejection fraction ≤35% were enrolled. One month after implantation, patients underwent a polysomnographic study. We evaluated the accuracy of the RDI for the prediction of severe SA (apnea-hypopnea index [AHI] ≥30 episodes/h) and the agreement between RDI and AHI during the sleep study night. Results: Two hundred sixty-five patients were enrolled to obtain the required sample of 173 patients with AHI and RDI data for analysis. The mean AHI was 21 ± 15 episodes/h and severe SA was diagnosed in 38 patients (22%), while the mean RDI was 33 ± 13 episodes/h. On the basis of the receiver operating characteristic curve analysis of RDI values, the area under the curve was 0.77 (95% confidence interval [CI] 0.70–0.83; P <.001). At an RDI value of 31 episodes/h, severe SA was detected with 87% (95% CI 72%–96%) sensitivity and 56% (95% CI 48%–66%) specificity. RDI closely correlated with AHI recorded during the same night (r = 0.74; 95% CI 0.57–0.84; P <.001), and the Bland-Altman agreement analysis revealed a bias of 11 episodes/h, with limits of agreement being −10 to 32 episodes/h. Conclusion: The RDI accurately identified severe SA and demonstrated good agreement with AHI. Therefore, it may serve as an efficient tool for screening patients at risk of SA.

Implantable cardioverter-defibrillator–computed respiratory disturbance index accurately identifies severe sleep apnea: The DASAP-HF study / D'Onofrio, Antonio; La Rovere, Maria Teresa; Emdin, Michele; Capucci, Alessandro; Sinagra, Gianfranco; Bianchi, Valter; Pisanò, Ennio C. L.; Pieragnoli, Paolo; Tespili, Maurizio; Luzi, Mario; Talarico, Antonello; Zecchin, Massimo; Rapacciuolo, Antonio; Piacenti, Marcello; Indolfi, Ciro; Arias, Miguel Angel; Diemberger, Igor; Checchinato, Catia; Boriani, Giuseppe; Padeletti, Luigi. - In: HEART RHYTHM. - ISSN 1547-5271. - 15:2(2018), pp. 211-217. [10.1016/j.hrthm.2017.09.038]

Implantable cardioverter-defibrillator–computed respiratory disturbance index accurately identifies severe sleep apnea: The DASAP-HF study

Boriani, Giuseppe;
2018

Abstract

Background: Sleep apnea (SA) is a relevant issue in the management of patients with heart failure for risk stratification and for implementing treatment strategies. Objective: The purpose of this study was to evaluate in patients with implantable cardioverter-defibrillators (ICDs) the performance of the respiratory disturbance index (RDI) computed by the ApneaScan algorithm (Boston Scientific Inc., Natick, MA) as a discriminator of severe SA. Methods: ICD-indicated patients with left ventricular ejection fraction ≤35% were enrolled. One month after implantation, patients underwent a polysomnographic study. We evaluated the accuracy of the RDI for the prediction of severe SA (apnea-hypopnea index [AHI] ≥30 episodes/h) and the agreement between RDI and AHI during the sleep study night. Results: Two hundred sixty-five patients were enrolled to obtain the required sample of 173 patients with AHI and RDI data for analysis. The mean AHI was 21 ± 15 episodes/h and severe SA was diagnosed in 38 patients (22%), while the mean RDI was 33 ± 13 episodes/h. On the basis of the receiver operating characteristic curve analysis of RDI values, the area under the curve was 0.77 (95% confidence interval [CI] 0.70–0.83; P <.001). At an RDI value of 31 episodes/h, severe SA was detected with 87% (95% CI 72%–96%) sensitivity and 56% (95% CI 48%–66%) specificity. RDI closely correlated with AHI recorded during the same night (r = 0.74; 95% CI 0.57–0.84; P <.001), and the Bland-Altman agreement analysis revealed a bias of 11 episodes/h, with limits of agreement being −10 to 32 episodes/h. Conclusion: The RDI accurately identified severe SA and demonstrated good agreement with AHI. Therefore, it may serve as an efficient tool for screening patients at risk of SA.
2018
7-ott-2017
15
2
211
217
Implantable cardioverter-defibrillator–computed respiratory disturbance index accurately identifies severe sleep apnea: The DASAP-HF study / D'Onofrio, Antonio; La Rovere, Maria Teresa; Emdin, Michele; Capucci, Alessandro; Sinagra, Gianfranco; Bianchi, Valter; Pisanò, Ennio C. L.; Pieragnoli, Paolo; Tespili, Maurizio; Luzi, Mario; Talarico, Antonello; Zecchin, Massimo; Rapacciuolo, Antonio; Piacenti, Marcello; Indolfi, Ciro; Arias, Miguel Angel; Diemberger, Igor; Checchinato, Catia; Boriani, Giuseppe; Padeletti, Luigi. - In: HEART RHYTHM. - ISSN 1547-5271. - 15:2(2018), pp. 211-217. [10.1016/j.hrthm.2017.09.038]
D'Onofrio, Antonio; La Rovere, Maria Teresa; Emdin, Michele; Capucci, Alessandro; Sinagra, Gianfranco; Bianchi, Valter; Pisanò, Ennio C. L.; Pieragnoli, Paolo; Tespili, Maurizio; Luzi, Mario; Talarico, Antonello; Zecchin, Massimo; Rapacciuolo, Antonio; Piacenti, Marcello; Indolfi, Ciro; Arias, Miguel Angel; Diemberger, Igor; Checchinato, Catia; Boriani, Giuseppe; Padeletti, Luigi
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S1547527117311359-main.pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 787.79 kB
Formato Adobe PDF
787.79 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1168118
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 16
social impact