Background: The rate of device replacement in pacemaker recipients has not been investigated in detail. Hypothesis: Current pacemakers with automatic management of atrial and ventricular pacing output provide sufficient longevity to minimize replacement rate. Methods: We considered a cohort of 542 pacemaker patients (age 78 ± 9 years, 60% male, 71% de-novo implants) and combined 1-month projected device longevity with survival data and late complication rate in a 3-state Markov model tested in several Monte Carlo computer simulations. Predetermined subgroups were: age < or ≥ 70; gender; primary indication to cardiac pacing. Results: At the 1-month follow-up the reported projected device longevity was 153 ± 45 months. With these values the proportion of patients expected to undergo a device replacement due to battery depletion was higher in patients aged <70 (49.9%, range 32.1%-61.9%) than in age ≥70 (24.5%, range 19.9%-28.8%); in women (39.9%, range 30.8%-48.1%) than in men (32.0%, range 24.7%-37.5%); in sinus node dysfunction (41.5%, range 30.2%-53.0%) than in atrio-ventricular block (33.5%, range 27.1-38.8%) or atrial fibrillation with bradycardia (27.9%, range 18.5%-37.0%). The expected replacement rate was inversely related to the assumed device longevity and depended on age class: a 50% increase in battery longevity implied a 5% reduction of replacement rates in patients aged ≥80. Conclusions: With current device technology 1/4 of pacemaker recipients aged ≥70 are expected to receive a second device in their life. Replacement rate depends on age, gender, and primary indication owing to differences in patients' survival expectancy. Additional improvements in device service time may modestly impact expected replacement rates especially in patients ≥80 years.
Impact of pacemaker longevity on expected device replacement rates: Results from computer simulations based on a multicenter registry (ESSENTIAL) / Boriani, Giuseppe; Bertini, Matteo; Saporito, Davide; Belotti, Giuseppina; Quartieri, Fabio; Tomasi, Corrado; Pucci, Angelo; Boggian, Giulio; Mazzocca, Gian Franco; Giorgi, Davide; Diotallevi, Paolo; Sassone, Biagio; Grassini, Diego; Gargaro, Alessio; Biffi, Mauro. - In: CLINICAL CARDIOLOGY. - ISSN 0160-9289. - 41:9(2018), pp. 1185-1191. [10.1002/clc.23003]
Impact of pacemaker longevity on expected device replacement rates: Results from computer simulations based on a multicenter registry (ESSENTIAL)
Boriani, Giuseppe;
2018
Abstract
Background: The rate of device replacement in pacemaker recipients has not been investigated in detail. Hypothesis: Current pacemakers with automatic management of atrial and ventricular pacing output provide sufficient longevity to minimize replacement rate. Methods: We considered a cohort of 542 pacemaker patients (age 78 ± 9 years, 60% male, 71% de-novo implants) and combined 1-month projected device longevity with survival data and late complication rate in a 3-state Markov model tested in several Monte Carlo computer simulations. Predetermined subgroups were: age < or ≥ 70; gender; primary indication to cardiac pacing. Results: At the 1-month follow-up the reported projected device longevity was 153 ± 45 months. With these values the proportion of patients expected to undergo a device replacement due to battery depletion was higher in patients aged <70 (49.9%, range 32.1%-61.9%) than in age ≥70 (24.5%, range 19.9%-28.8%); in women (39.9%, range 30.8%-48.1%) than in men (32.0%, range 24.7%-37.5%); in sinus node dysfunction (41.5%, range 30.2%-53.0%) than in atrio-ventricular block (33.5%, range 27.1-38.8%) or atrial fibrillation with bradycardia (27.9%, range 18.5%-37.0%). The expected replacement rate was inversely related to the assumed device longevity and depended on age class: a 50% increase in battery longevity implied a 5% reduction of replacement rates in patients aged ≥80. Conclusions: With current device technology 1/4 of pacemaker recipients aged ≥70 are expected to receive a second device in their life. Replacement rate depends on age, gender, and primary indication owing to differences in patients' survival expectancy. Additional improvements in device service time may modestly impact expected replacement rates especially in patients ≥80 years.File | Dimensione | Formato | |
---|---|---|---|
clc.23003.pdf
Accesso riservato
Tipologia:
Versione pubblicata dall'editore
Dimensione
1.2 MB
Formato
Adobe PDF
|
1.2 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
clc.23003.pdf
Open access
Tipologia:
Versione dell'autore revisionata e accettata per la pubblicazione
Dimensione
576.14 kB
Formato
Adobe PDF
|
576.14 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris