Aims We evaluated clinical and radiographic outcomes of total shoulder arthroplasty (TSA) using the second-generation Trabecular Metal (TM) Glenoid component. The first generation component was withdrawn in 2005 after a series of failures were reported. Between 2009 and 2012, 40 consecutive patients with unilateral TSA using the second-generation component were enrolled in this clinical study. The mean age of the patients was 63.8 years (40 to 75) and the mean follow-up was 38 months (24 to 42). Methods Patients were evaluated using the Constant score (CS), the American Shoulder and Elbow Surg.s (ASES) score and routine radiographs. Results Significant differences were found between the pre-And post-operative CS (p = 0.003), ASES (p = 0.009) scores and CS subscores of pain (p < 0.001), strength (p < 0.001) and mobility items (p < 0.05). No glenoid or humeral components migrated. Posterior thinning of the keel and slight wear at the polyethylene-TM interface was observed in one patient but was asymptomatic. Radiolucent lines were found around three humeral (< 1.5 mm) and two glenoid components (< 1 mm) and all were asymptomatic. Discussion TSA with the second-generation TM Glenoid component results in satisfactory to excellent clinical performance, function, and subjective satisfaction at a mean follow-up of about three years. Radiographic changes were few and did not affect the outcome. Take home message: This paper highlights that the second generation Trabecular Metal Glenoid has better outcomes than those reported with the first-generation component.

Total shoulder arthroplasty with a secondgeneration tantalum trabecular metal-backed glenoid component / Merolla, G.; Chin, P.; Sasyniuk, T. M.; Paladini, P.; Porcellini, G.. - In: THE BONE & JOINT JOURNAL. - ISSN 2049-4394. - 98B:1(2016), pp. 75-80. [10.1302/0301-620X.98B1.36620]

Total shoulder arthroplasty with a secondgeneration tantalum trabecular metal-backed glenoid component

Porcellini, G.
2016

Abstract

Aims We evaluated clinical and radiographic outcomes of total shoulder arthroplasty (TSA) using the second-generation Trabecular Metal (TM) Glenoid component. The first generation component was withdrawn in 2005 after a series of failures were reported. Between 2009 and 2012, 40 consecutive patients with unilateral TSA using the second-generation component were enrolled in this clinical study. The mean age of the patients was 63.8 years (40 to 75) and the mean follow-up was 38 months (24 to 42). Methods Patients were evaluated using the Constant score (CS), the American Shoulder and Elbow Surg.s (ASES) score and routine radiographs. Results Significant differences were found between the pre-And post-operative CS (p = 0.003), ASES (p = 0.009) scores and CS subscores of pain (p < 0.001), strength (p < 0.001) and mobility items (p < 0.05). No glenoid or humeral components migrated. Posterior thinning of the keel and slight wear at the polyethylene-TM interface was observed in one patient but was asymptomatic. Radiolucent lines were found around three humeral (< 1.5 mm) and two glenoid components (< 1 mm) and all were asymptomatic. Discussion TSA with the second-generation TM Glenoid component results in satisfactory to excellent clinical performance, function, and subjective satisfaction at a mean follow-up of about three years. Radiographic changes were few and did not affect the outcome. Take home message: This paper highlights that the second generation Trabecular Metal Glenoid has better outcomes than those reported with the first-generation component.
2016
98B
1
75
80
Total shoulder arthroplasty with a secondgeneration tantalum trabecular metal-backed glenoid component / Merolla, G.; Chin, P.; Sasyniuk, T. M.; Paladini, P.; Porcellini, G.. - In: THE BONE & JOINT JOURNAL. - ISSN 2049-4394. - 98B:1(2016), pp. 75-80. [10.1302/0301-620X.98B1.36620]
Merolla, G.; Chin, P.; Sasyniuk, T. M.; Paladini, P.; Porcellini, G.
File in questo prodotto:
File Dimensione Formato  
0301-620x.98b1.36620.pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 1.27 MB
Formato Adobe PDF
1.27 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1160303
Citazioni
  • ???jsp.display-item.citation.pmc??? 16
  • Scopus 36
  • ???jsp.display-item.citation.isi??? 30
social impact