Baseline wander is a kind of noise that affects all ECG signals. We have recently proposed a novel approach to its removal which is based on Quadratic Variation Reduction (QVR). The approach has very favorable properties and has shown to be effective in removing baseline wander, while preserving the ST segment level. It requires the determination of a detrending parameter. In this pa- per, we derive a linear time-invariant filter approximating QVR. The filter retains (approximately) the same optimality properties as QVR. Moreover, it provides a criterion for choosing the proper value of the parameter governing QVR, as a function of the spectral characteristics of base- line wander noise. Simulation results show that the filter is effective in removing baseline wander, while introducing minor distortion in the ST segment.
Baseline Wander Removal in ECG and AHA Recommendations / Fasano, Antonio; Villani, Valeria. - In: COMPUTING IN CARDIOLOGY. - ISSN 2325-8861. - 40:(2013), pp. 1171-1174. (Intervento presentato al convegno 2013 40th Computing in Cardiology Conference, CinC 2013 tenutosi a Zaragoza, esp nel Sep. 23-25, 2013).
Baseline Wander Removal in ECG and AHA Recommendations
Villani, Valeria
2013
Abstract
Baseline wander is a kind of noise that affects all ECG signals. We have recently proposed a novel approach to its removal which is based on Quadratic Variation Reduction (QVR). The approach has very favorable properties and has shown to be effective in removing baseline wander, while preserving the ST segment level. It requires the determination of a detrending parameter. In this pa- per, we derive a linear time-invariant filter approximating QVR. The filter retains (approximately) the same optimality properties as QVR. Moreover, it provides a criterion for choosing the proper value of the parameter governing QVR, as a function of the spectral characteristics of base- line wander noise. Simulation results show that the filter is effective in removing baseline wander, while introducing minor distortion in the ST segment.File | Dimensione | Formato | |
---|---|---|---|
Baseline Wander Removal in ECG and AHA Recommendations.pdf
Accesso riservato
Tipologia:
VOR - Versione pubblicata dall'editore
Dimensione
694.08 kB
Formato
Adobe PDF
|
694.08 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris