Aims Implantable cardioverter defibrillators improve survival of patients at risk for ventricular arrhythmias, but inappropriate shocks occur in up to 30% of patients and have been associated with worse quality of life and prognosis. In heart failure patients with cardiac resynchronization therapy defibrillators (CRT-Ds), we evaluated whether a new generation of detection and discrimination algorithms reduces inappropriate shocks. Methods and results We analysed 1983 Medtronic CRT-D patients (80% male, 67±10 years), 1368 with standard devices (Control CRTD) and 615 with new generation devices (New CRT-D). Expert electrophysiologists reviewed and classified the electrograms of all device-detected ventricular tachycardia/fibrillation episodes. Total follow-up was 3751 patients-years. Incidence of inappropriate shocks at 1 year was 2.8% [95% confidence interval (CI) = 2.0-3.5] in Control CRT-D and 0.9% (CI = 0.4-2.2) in New CRT-D (hazard ratio = 0.37, CI = 0.21-0.66, P < 0.001). In New CRT-D, inappropriate shocks were reduced by 77% [incidence rate ratio (IRR) = 0.23, CI = 0.16-0.35, P < 0.001] and inappropriate antitachycardia pacing by 81% (IRR = 0.19, CI = 0.11-0.335, P < 0.001). Annual rate per 100 patient-years for appropriate VF detections was 3.0 (CI = 2.1-4.2) in New CRT-D and 3.2 (CI = 2.1-5.0) in Control CRT-D (P = 0.68), for syncope was 0.4 (CI = 0.2-0.9) in New CRT-D and 0.7 (CI = 0.5-1.0) in Control CRT-D (P = 0.266), and for death was 1.0 (CI = 0.6-1.6) in New CRT-D and 3.5 (CI = 3.0-4.1) in Control CRT-D (P < 0.001). Conclusion Detection and discrimination algorithms used in new generation CRT-D significantly reduced inappropriate shocks when compared with standard CRT-D. This result, with no compromise on VF sensitivity or risk of syncope, has important implications for patients' quality of life and prognosis.

Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: A historical comparison of a prospective database / Lunati, Maurizio; Proclemer, Alessandro; Boriani, Giuseppe; Landolina, Maurizio; Locati, Emanuela; Rordorf, Roberto; Daleffe, Elisabetta; Ricci, Renato Pietro; Catanzariti, Domenico; Tomasi, Luca; Gulizia, Michele; Baccillieri, Maria Stella; Molon, Giulio; Gasparini, Maurizio. - In: EUROPACE. - ISSN 1099-5129. - 18:9(2016), pp. 1391-1398. [10.1093/europace/euv420]

Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: A historical comparison of a prospective database

BORIANI, Giuseppe;
2016

Abstract

Aims Implantable cardioverter defibrillators improve survival of patients at risk for ventricular arrhythmias, but inappropriate shocks occur in up to 30% of patients and have been associated with worse quality of life and prognosis. In heart failure patients with cardiac resynchronization therapy defibrillators (CRT-Ds), we evaluated whether a new generation of detection and discrimination algorithms reduces inappropriate shocks. Methods and results We analysed 1983 Medtronic CRT-D patients (80% male, 67±10 years), 1368 with standard devices (Control CRTD) and 615 with new generation devices (New CRT-D). Expert electrophysiologists reviewed and classified the electrograms of all device-detected ventricular tachycardia/fibrillation episodes. Total follow-up was 3751 patients-years. Incidence of inappropriate shocks at 1 year was 2.8% [95% confidence interval (CI) = 2.0-3.5] in Control CRT-D and 0.9% (CI = 0.4-2.2) in New CRT-D (hazard ratio = 0.37, CI = 0.21-0.66, P < 0.001). In New CRT-D, inappropriate shocks were reduced by 77% [incidence rate ratio (IRR) = 0.23, CI = 0.16-0.35, P < 0.001] and inappropriate antitachycardia pacing by 81% (IRR = 0.19, CI = 0.11-0.335, P < 0.001). Annual rate per 100 patient-years for appropriate VF detections was 3.0 (CI = 2.1-4.2) in New CRT-D and 3.2 (CI = 2.1-5.0) in Control CRT-D (P = 0.68), for syncope was 0.4 (CI = 0.2-0.9) in New CRT-D and 0.7 (CI = 0.5-1.0) in Control CRT-D (P = 0.266), and for death was 1.0 (CI = 0.6-1.6) in New CRT-D and 3.5 (CI = 3.0-4.1) in Control CRT-D (P < 0.001). Conclusion Detection and discrimination algorithms used in new generation CRT-D significantly reduced inappropriate shocks when compared with standard CRT-D. This result, with no compromise on VF sensitivity or risk of syncope, has important implications for patients' quality of life and prognosis.
2016
29-gen-2016
18
9
1391
1398
Reduction of inappropriate anti-tachycardia pacing therapies and shocks by a novel suite of detection algorithms in heart failure patients with cardiac resynchronization therapy defibrillators: A historical comparison of a prospective database / Lunati, Maurizio; Proclemer, Alessandro; Boriani, Giuseppe; Landolina, Maurizio; Locati, Emanuela; Rordorf, Roberto; Daleffe, Elisabetta; Ricci, Renato Pietro; Catanzariti, Domenico; Tomasi, Luca; Gulizia, Michele; Baccillieri, Maria Stella; Molon, Giulio; Gasparini, Maurizio. - In: EUROPACE. - ISSN 1099-5129. - 18:9(2016), pp. 1391-1398. [10.1093/europace/euv420]
Lunati, Maurizio; Proclemer, Alessandro; Boriani, Giuseppe; Landolina, Maurizio; Locati, Emanuela; Rordorf, Roberto; Daleffe, Elisabetta; Ricci, Renato Pietro; Catanzariti, Domenico; Tomasi, Luca; Gulizia, Michele; Baccillieri, Maria Stella; Molon, Giulio; Gasparini, Maurizio
File in questo prodotto:
File Dimensione Formato  
euv420.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 204.81 kB
Formato Adobe PDF
204.81 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1139568
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 2
social impact