Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.
Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD / Corzani, Alessandro; Ziacchi, Matteo; Biffi, Mauro; Diemberger, Igor; Martignani, Cristian; Boriani, Giuseppe. - In: INDIAN HEART JOURNAL. - ISSN 0019-4832. - 67:1(2015), pp. 56-59. [10.1016/j.ihj.2015.01.001]
Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD
BORIANI, Giuseppe
2015
Abstract
Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S0019483215000048-main.pdf
Open access
Tipologia:
Versione pubblicata dall'editore
Dimensione
885.56 kB
Formato
Adobe PDF
|
885.56 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris