Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.

Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD / Corzani, Alessandro; Ziacchi, Matteo; Biffi, Mauro; Diemberger, Igor; Martignani, Cristian; Boriani, Giuseppe. - In: INDIAN HEART JOURNAL. - ISSN 0019-4832. - 67:1(2015), pp. 56-59. [10.1016/j.ihj.2015.01.001]

Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD

BORIANI, Giuseppe
2015

Abstract

Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.
2015
12-feb-2015
67
1
56
59
Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD / Corzani, Alessandro; Ziacchi, Matteo; Biffi, Mauro; Diemberger, Igor; Martignani, Cristian; Boriani, Giuseppe. - In: INDIAN HEART JOURNAL. - ISSN 0019-4832. - 67:1(2015), pp. 56-59. [10.1016/j.ihj.2015.01.001]
Corzani, Alessandro; Ziacchi, Matteo; Biffi, Mauro; Diemberger, Igor; Martignani, Cristian; Boriani, Giuseppe
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S0019483215000048-main.pdf

Open access

Tipologia: Versione pubblicata dall'editore
Dimensione 885.56 kB
Formato Adobe PDF
885.56 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1139265
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 13
  • ???jsp.display-item.citation.isi??? ND
social impact