Autocapture is an algorithm for automatic adaptation of ventricular output to capture threshold. The aim of this prospective study was to estimate the effects of ventricular Autocapture algorithm on DDD-DDDR pacemaker longevity. METHODS AND RESULTS: Eighty-three patients implanted with a DDD-DDDR pacemaker (Affinity or Entity; St Jude Medical, USA) were enrolled and the Autocapture function was activated pre-discharge. Ventricular pulse duration was randomly programmed at 0.3 or 0.4 ms, with a cross-over at 8-12 weeks and again at 13-14 months. Diagnostic data were retrieved from device memory and by calculating battery current drain from long-term threshold recordings; device longevity was estimated at the following settings: Autocapture with a pulse duration of 0.3 and 0.4 ms, respectively, standard output (3.5 V, 0.4 ms) and conventional low output programming (2.5 V, 0.4 ms). According to a series of assumptions, Autocapture was associated with a 55-60% increase in estimated device longevity compared with standard output programming and a 6-7% increase in longevity compared with low output programming. No significant differences were found between Autocapture programmed with a pulse duration of 0.3 or 0.4 ms. In projections to a 10-year follow-up, use of the Autocapture function resulted in a 42% reduction in pacing-related estimated costs compared with standard output programming at 3.5 V, 0.4 ms. CONCLUSION: Pacing with constant adaptation of ventricular output in dual-chamber devices has the potential to increase generator longevity and to reduce sizeably pacing-related costs compared with standard programming

Role of ventricular Autocapture function in increasing longevity of DDDR pacemakers: a prospective study / Boriani, Giuseppe; Rusconi, L; Biffi, M; Pavia, L; Sassara, M; Malfitano, D; Bongiorni, Mg; Padeletti, L; Filice, I; Sanfelici, D; Maffei, P; Vicentini, A; Branzi, A.. - In: EUROPACE. - ISSN 1099-5129. - 8:3(2006), pp. 216-220. [10.1093/europace/euj027]

Role of ventricular Autocapture function in increasing longevity of DDDR pacemakers: a prospective study.

BORIANI, Giuseppe;
2006

Abstract

Autocapture is an algorithm for automatic adaptation of ventricular output to capture threshold. The aim of this prospective study was to estimate the effects of ventricular Autocapture algorithm on DDD-DDDR pacemaker longevity. METHODS AND RESULTS: Eighty-three patients implanted with a DDD-DDDR pacemaker (Affinity or Entity; St Jude Medical, USA) were enrolled and the Autocapture function was activated pre-discharge. Ventricular pulse duration was randomly programmed at 0.3 or 0.4 ms, with a cross-over at 8-12 weeks and again at 13-14 months. Diagnostic data were retrieved from device memory and by calculating battery current drain from long-term threshold recordings; device longevity was estimated at the following settings: Autocapture with a pulse duration of 0.3 and 0.4 ms, respectively, standard output (3.5 V, 0.4 ms) and conventional low output programming (2.5 V, 0.4 ms). According to a series of assumptions, Autocapture was associated with a 55-60% increase in estimated device longevity compared with standard output programming and a 6-7% increase in longevity compared with low output programming. No significant differences were found between Autocapture programmed with a pulse duration of 0.3 or 0.4 ms. In projections to a 10-year follow-up, use of the Autocapture function resulted in a 42% reduction in pacing-related estimated costs compared with standard output programming at 3.5 V, 0.4 ms. CONCLUSION: Pacing with constant adaptation of ventricular output in dual-chamber devices has the potential to increase generator longevity and to reduce sizeably pacing-related costs compared with standard programming
2006
8
3
216
220
Role of ventricular Autocapture function in increasing longevity of DDDR pacemakers: a prospective study / Boriani, Giuseppe; Rusconi, L; Biffi, M; Pavia, L; Sassara, M; Malfitano, D; Bongiorni, Mg; Padeletti, L; Filice, I; Sanfelici, D; Maffei, P; Vicentini, A; Branzi, A.. - In: EUROPACE. - ISSN 1099-5129. - 8:3(2006), pp. 216-220. [10.1093/europace/euj027]
Boriani, Giuseppe; Rusconi, L; Biffi, M; Pavia, L; Sassara, M; Malfitano, D; Bongiorni, Mg; Padeletti, L; Filice, I; Sanfelici, D; Maffei, P; Vicentin...espandi
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1080432
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 50
  • ???jsp.display-item.citation.isi??? 44
social impact