Background: Single-lead VDD pacing systems are an alternative to conventional DDD pacemakers in patients with atrioventricular (AV) block and normal sinus function. Hypothesis: The aim of this study was to assess changes of P-wave amplitude occurring in dynamic conditions in two groups of patients with a single-lead VDD and with a DDD pacing system, respectively. Methods: Twenty-eight patients with second- or third-degree AV block and normal sinus function were enrolled prospectively into the study. Seventeen patients were implanted with a single-lead VDD pacing system and 11 with a DDD pacemaker. Patients were evaluated at 3 months tall patients) and at 6 months (26 patients) at supine and in dynamic conditions (postural changes, hyperventilation, and during exercise). Results: Mean P-wave values at supine were 1.92 +/- 1.10 mV at 3 months and 1.76 +/- 1.01 mV at 6 months for VDD systems, and 4.63 +/- 2.18 mV at 3 months and 4.58 +/- 2.80 mV at 6 months for DDD pacemakers. In dynamic conditions, P-wave amplitude changes compared with supine condition ranged between -74 and +226% in VDD, and between -53 and +138% in DDD; however P-wave amplitudes showed no significant changes compared with baseline. Moreover, changes in atrial signal amplitudes did not occur randomly, and in both systems P-wave amplitudes remained significantly correlated with supine values. Conclusions: A wide range of P-wave amplitude variations occurs in different postural conditions or during exercise, both with single-lead VDD and DDD pacing systems. However, with appropriate programming of atrial sensitivity based on supine values, constant atrial tracking can be maintained.
DDD and single-lead VDD pacing: Evaluation of atrial signal dynamic changes / Boriani, Giuseppe; Biffi, M; Bandini, A; Moracchini, Pv; Capucci, A; Sigliano, R; Mininno, A; Branzi, A.. - In: CLINICAL CARDIOLOGY. - ISSN 0160-9289. - 23:9(2000), pp. 678-680. [10.1002/clc.4960230909]
DDD and single-lead VDD pacing: Evaluation of atrial signal dynamic changes
BORIANI, Giuseppe;
2000
Abstract
Background: Single-lead VDD pacing systems are an alternative to conventional DDD pacemakers in patients with atrioventricular (AV) block and normal sinus function. Hypothesis: The aim of this study was to assess changes of P-wave amplitude occurring in dynamic conditions in two groups of patients with a single-lead VDD and with a DDD pacing system, respectively. Methods: Twenty-eight patients with second- or third-degree AV block and normal sinus function were enrolled prospectively into the study. Seventeen patients were implanted with a single-lead VDD pacing system and 11 with a DDD pacemaker. Patients were evaluated at 3 months tall patients) and at 6 months (26 patients) at supine and in dynamic conditions (postural changes, hyperventilation, and during exercise). Results: Mean P-wave values at supine were 1.92 +/- 1.10 mV at 3 months and 1.76 +/- 1.01 mV at 6 months for VDD systems, and 4.63 +/- 2.18 mV at 3 months and 4.58 +/- 2.80 mV at 6 months for DDD pacemakers. In dynamic conditions, P-wave amplitude changes compared with supine condition ranged between -74 and +226% in VDD, and between -53 and +138% in DDD; however P-wave amplitudes showed no significant changes compared with baseline. Moreover, changes in atrial signal amplitudes did not occur randomly, and in both systems P-wave amplitudes remained significantly correlated with supine values. Conclusions: A wide range of P-wave amplitude variations occurs in different postural conditions or during exercise, both with single-lead VDD and DDD pacing systems. However, with appropriate programming of atrial sensitivity based on supine values, constant atrial tracking can be maintained.File | Dimensione | Formato | |
---|---|---|---|
Clinical Cardiology - 2009 - Boriani - DDD and single‐lead VDD pacing Evaluation of atrial signal dynamic changes.pdf
Open access
Tipologia:
VOR - Versione pubblicata dall'editore
Dimensione
320.7 kB
Formato
Adobe PDF
|
320.7 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris