Background: We studied long-term right ventricular (RV) pacing threshold (RVPT) behavior in patients consecutively implanted with pacemakers capable of automatic output reprogramming tracked by automatic RV threshold measurement (automatic verification of capture [AVC]). Methods: All the patients had state-of-the art steroid-eluting bipolar pacing leads and were RV-paced by an AVC algorithm from the three American manufacturers. Follow-up occurred twice in the first year after implantation, then yearly until approaching elective replacement indicator. Results: Three hundred and twenty-one patients aged 73 ± 12 years were observed for 49 ± 26 months on average. At implantation, RVPT was 0.54 ± 0.2 V at 0.4 ms at an average 774 ± 217 Ω impedance. Forty-one of the 321 patients (12.8%) had a permanent RVPT increase above 1.5 V at 0.4 ms: RVPT was between 1.6 and 2.5 V in 29 of 321 (9%) patients, whereas it was between 2.6 and 3.5 V in seven of 321 (2.2%) patients, and >3.5 V in five of 321 (1.5%) patients. No exit block occurred because of automatic RV output adjustment by AVC algorithms. No predictor of RVPT increase was found at multivariable analysis. The maximum RVPT increase occurred within 12 months from implantation in 19 of 321 (5.9%) patients, between the first and the second year in 12 of 321 (3.7%), between the second and the sixth year in eight of 321 (2.5%), and after the sixth year in two of 321 (0.6%). Conclusion: Despite technologic improvement in lead manufacturing, long-term increase of the RVPT occurs in about 13% of patients, possibly representing a serious safety issue in 3.7% when 2.5 V at 0.4 ms is exceeded. AVC algorithms can improve patients' safety by automatic tailoring of the pacing output to threshold fluctuations, while maximizing device longevity

Long-term RV threshold behavior by automated measurements: safety is the standpoint of pacemaker longevity! / Biffi, M; Bertini, M; Mazzotti, A; Gardini, B; Mantovani, V; Ziacchi, M; Valzania, C; Martignani, C; Diemberger, I; Boriani, Giuseppe. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 34:(2011), pp. 89-95. [10.1111/j.1540-8159.2010.02915.x]

Long-term RV threshold behavior by automated measurements: safety is the standpoint of pacemaker longevity!

BORIANI, Giuseppe
2011

Abstract

Background: We studied long-term right ventricular (RV) pacing threshold (RVPT) behavior in patients consecutively implanted with pacemakers capable of automatic output reprogramming tracked by automatic RV threshold measurement (automatic verification of capture [AVC]). Methods: All the patients had state-of-the art steroid-eluting bipolar pacing leads and were RV-paced by an AVC algorithm from the three American manufacturers. Follow-up occurred twice in the first year after implantation, then yearly until approaching elective replacement indicator. Results: Three hundred and twenty-one patients aged 73 ± 12 years were observed for 49 ± 26 months on average. At implantation, RVPT was 0.54 ± 0.2 V at 0.4 ms at an average 774 ± 217 Ω impedance. Forty-one of the 321 patients (12.8%) had a permanent RVPT increase above 1.5 V at 0.4 ms: RVPT was between 1.6 and 2.5 V in 29 of 321 (9%) patients, whereas it was between 2.6 and 3.5 V in seven of 321 (2.2%) patients, and >3.5 V in five of 321 (1.5%) patients. No exit block occurred because of automatic RV output adjustment by AVC algorithms. No predictor of RVPT increase was found at multivariable analysis. The maximum RVPT increase occurred within 12 months from implantation in 19 of 321 (5.9%) patients, between the first and the second year in 12 of 321 (3.7%), between the second and the sixth year in eight of 321 (2.5%), and after the sixth year in two of 321 (0.6%). Conclusion: Despite technologic improvement in lead manufacturing, long-term increase of the RVPT occurs in about 13% of patients, possibly representing a serious safety issue in 3.7% when 2.5 V at 0.4 ms is exceeded. AVC algorithms can improve patients' safety by automatic tailoring of the pacing output to threshold fluctuations, while maximizing device longevity
2011
34
89
95
Long-term RV threshold behavior by automated measurements: safety is the standpoint of pacemaker longevity! / Biffi, M; Bertini, M; Mazzotti, A; Gardini, B; Mantovani, V; Ziacchi, M; Valzania, C; Martignani, C; Diemberger, I; Boriani, Giuseppe. - In: PACING AND CLINICAL ELECTROPHYSIOLOGY. - ISSN 0147-8389. - 34:(2011), pp. 89-95. [10.1111/j.1540-8159.2010.02915.x]
Biffi, M; Bertini, M; Mazzotti, A; Gardini, B; Mantovani, V; Ziacchi, M; Valzania, C; Martignani, C; Diemberger, I; Boriani, Giuseppe
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1080144
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 22
  • ???jsp.display-item.citation.isi??? 19
social impact