A simple measuring system for early detection of haemolysis during haemodialysis is proposed. The developed measuring system performs an optical absorption spectroscopy analysis of the dialysis fluid in order to estimate the free haemoglobin concentration in the blood returned to the patient-the potential cause of injury for the patient. The integration of the developed measuring system into hemodialyzers results simple and relatively inexpensive (overall system cost is about few hundred dollars). Preliminary results confirm the applicability of the proposed measuring system.

A simple measuring system for early detection of haemolysis during haemodialysis / Cattini, Stefano; Rovati, Luigi; M., Bernabei; D., Cianciavicchia; P., Monari; A., Sicuri. - STAMPA. - 1:(2014), pp. 282-287. (Intervento presentato al convegno 2014 IEEE International Instrumentation and Measurement Technology Conference: Instrumentation and Measurement for Sustainable Development, I2MTC 2014 tenutosi a Montevideo, ury nel 12-15 May 2014) [10.1109/I2MTC.2014.6860753].

A simple measuring system for early detection of haemolysis during haemodialysis

CATTINI, Stefano;ROVATI, Luigi;
2014

Abstract

A simple measuring system for early detection of haemolysis during haemodialysis is proposed. The developed measuring system performs an optical absorption spectroscopy analysis of the dialysis fluid in order to estimate the free haemoglobin concentration in the blood returned to the patient-the potential cause of injury for the patient. The integration of the developed measuring system into hemodialyzers results simple and relatively inexpensive (overall system cost is about few hundred dollars). Preliminary results confirm the applicability of the proposed measuring system.
2014
2014 IEEE International Instrumentation and Measurement Technology Conference: Instrumentation and Measurement for Sustainable Development, I2MTC 2014
Montevideo, ury
12-15 May 2014
1
282
287
Cattini, Stefano; Rovati, Luigi; M., Bernabei; D., Cianciavicchia; P., Monari; A., Sicuri
A simple measuring system for early detection of haemolysis during haemodialysis / Cattini, Stefano; Rovati, Luigi; M., Bernabei; D., Cianciavicchia; P., Monari; A., Sicuri. - STAMPA. - 1:(2014), pp. 282-287. (Intervento presentato al convegno 2014 IEEE International Instrumentation and Measurement Technology Conference: Instrumentation and Measurement for Sustainable Development, I2MTC 2014 tenutosi a Montevideo, ury nel 12-15 May 2014) [10.1109/I2MTC.2014.6860753].
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1051324
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 9
social impact