The aim of this study was to test the usefulness of hygroscopic condenser humidifiers on secretion and on inspired gas temperature in tracheostomized patients. Forty spontaneously breathing chronically tracheostomized patients were divided into two groups: Group 1 received a hygroscopic condenser humidifier connected to the tracheostomy, 24 h daily for 10 days; Group 2, without any protection system, was chosen as the control group. The daily number of tracheal suctions, quantity of aspirate and thickness and colouring of secretions was evaluated. At baseline, and at days 5 and 10, patients were submitted to blood gas analysis, respiratory function tests and sputum analysis. The temperature of gases breathed was measured at rest and during a hyperventilation test, with and without the hygroscopic condenser humidifier. Statistically significant differences were found in thickness and colouring of secretions between the two groups during the period of 10 days. Group 2 showed a significantly greater trend in number of bacteria than Group 1. The group with the hygroscopic condenser humidifier showed respiratory function improvement over time for forced expiratory volume in one second (FEV1) and tidal volume (VT), maximal inspiratory pressure (MIP), and maximal voluntary ventilation (MVV) in comparison to the control group, who did not. Significant differences in the temperature between rest and hyperventilation, with and without a hygroscopic condenser humidifier were also found. In conclusion, a hygroscopic condenser humidifier may be useful in chronically tracheostomized patients who breathe spontaneously, improving viscosity and colouring of secretions, preventing further bacterial colonization, heating inspiratory flow, and helping to improve the functional outcome.

Hygroscopic condenser humidifiers in chronically tracheostomized patients who breathe spontaneously / M., Vitacca; Clini, Enrico; K., Foglio; S., Scalvini; S., Marangoni; A., Quadri; N., Ambrosino. - In: EUROPEAN RESPIRATORY JOURNAL. - ISSN 0903-1936. - ELETTRONICO. - 7:(1994), pp. 2026-2032.

Hygroscopic condenser humidifiers in chronically tracheostomized patients who breathe spontaneously.

CLINI, Enrico;
1994

Abstract

The aim of this study was to test the usefulness of hygroscopic condenser humidifiers on secretion and on inspired gas temperature in tracheostomized patients. Forty spontaneously breathing chronically tracheostomized patients were divided into two groups: Group 1 received a hygroscopic condenser humidifier connected to the tracheostomy, 24 h daily for 10 days; Group 2, without any protection system, was chosen as the control group. The daily number of tracheal suctions, quantity of aspirate and thickness and colouring of secretions was evaluated. At baseline, and at days 5 and 10, patients were submitted to blood gas analysis, respiratory function tests and sputum analysis. The temperature of gases breathed was measured at rest and during a hyperventilation test, with and without the hygroscopic condenser humidifier. Statistically significant differences were found in thickness and colouring of secretions between the two groups during the period of 10 days. Group 2 showed a significantly greater trend in number of bacteria than Group 1. The group with the hygroscopic condenser humidifier showed respiratory function improvement over time for forced expiratory volume in one second (FEV1) and tidal volume (VT), maximal inspiratory pressure (MIP), and maximal voluntary ventilation (MVV) in comparison to the control group, who did not. Significant differences in the temperature between rest and hyperventilation, with and without a hygroscopic condenser humidifier were also found. In conclusion, a hygroscopic condenser humidifier may be useful in chronically tracheostomized patients who breathe spontaneously, improving viscosity and colouring of secretions, preventing further bacterial colonization, heating inspiratory flow, and helping to improve the functional outcome.
1994
7
2026
2032
Hygroscopic condenser humidifiers in chronically tracheostomized patients who breathe spontaneously / M., Vitacca; Clini, Enrico; K., Foglio; S., Scalvini; S., Marangoni; A., Quadri; N., Ambrosino. - In: EUROPEAN RESPIRATORY JOURNAL. - ISSN 0903-1936. - ELETTRONICO. - 7:(1994), pp. 2026-2032.
M., Vitacca; Clini, Enrico; K., Foglio; S., Scalvini; S., Marangoni; A., Quadri; N., Ambrosino
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/593128
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
social impact