Lifestyle change targeted towards increasing daily resting energy expenditure (REE) is one of the cornerstones of obesity treatment. Measurements of energy expenditure and substrate utilization are essential to understanding the metabolic basis of obesity, and the physiological responses to perturbations in habitual food intake. REE is the largest part of human energy expenditure (60-70%) and an increase or decrease in REE would have a large impact on total energy. Accurate and easy-to-use methods for measuring REE are needed, to be applied by clinicians in daily clinical settings to assess the validity of a new instrument to estimate REE in normal weight, healthy adults. METHODS: Ninety-nine subjects (52 females and 47 males) (mean+/-SD, age 38+/-14 years; body mass index (BMI) 23+/-3 kg/m(2)) were tested. REE was assessed using a Sensor Medics Vmax metabolic cart with a ventilated canopy and with the SenseWear armband. Body composition, percentage fat mass (%FM) and percentage fat free mass (ÿM) were assessed by skinfold thickness measurements (SF), bio-electrical impedance analysis (BIA) and air displacement plethysmography (BOD-POD). RESULTS: No significant difference was found among measurements of FFM using the three different techniques. Both SenseWear and Sensor Medics Vmax showed a high correlation, r=0.42 and r=0.40 (p<0.0001) respectively, with BMI. No significant difference was found in mean REE between SenseWear (1540+/-280 kcal/day) and Sensor Medics Vmax (1700+/-330 kcal/day) (p=ns) and the correlation between REE measured by SenseWear and Sensor Medics Vmax was high (r=0.86, p<0.0001). Bland-Altman plot showed no difference in REE determination between SenseWear and Sensor Medics Vmax. ÿM determined by BOD-POD correlated with SenseWear (r=0.42, p<0.0001) as well as Sensor Medics Vmax (r=0.38, p<0.001). CONCLUSION: SF, BIA and BOD-POD provide valid and reliable measurements of FFM. Our results suggest that the SenseWear armband is an acceptable device to accurately measure REE in healthy subjects. Its characteristics have the potential to reduce measurement times and make the SenseWear armband useful for epidemiological studies.

A new device for measuring resting energy expenditure (REE) in healthy subjects / Malavolti, Marcella; A., Pietrobelli; Dugoni, Manfredo; M., Poli; Romagnoli, Elisa; Battistini, Nino Carlo. - In: NMCD. NUTRITION METABOLISM AND CARDIOVASCULAR DISEASES. - ISSN 0939-4753. - STAMPA. - 17:5(2007), pp. 338-343. [10.1016/j.numecd.2005.12.009]

A new device for measuring resting energy expenditure (REE) in healthy subjects.

MALAVOLTI, Marcella;DUGONI, Manfredo;ROMAGNOLI, Elisa;BATTISTINI, Nino Carlo
2007

Abstract

Lifestyle change targeted towards increasing daily resting energy expenditure (REE) is one of the cornerstones of obesity treatment. Measurements of energy expenditure and substrate utilization are essential to understanding the metabolic basis of obesity, and the physiological responses to perturbations in habitual food intake. REE is the largest part of human energy expenditure (60-70%) and an increase or decrease in REE would have a large impact on total energy. Accurate and easy-to-use methods for measuring REE are needed, to be applied by clinicians in daily clinical settings to assess the validity of a new instrument to estimate REE in normal weight, healthy adults. METHODS: Ninety-nine subjects (52 females and 47 males) (mean+/-SD, age 38+/-14 years; body mass index (BMI) 23+/-3 kg/m(2)) were tested. REE was assessed using a Sensor Medics Vmax metabolic cart with a ventilated canopy and with the SenseWear armband. Body composition, percentage fat mass (%FM) and percentage fat free mass (ÿM) were assessed by skinfold thickness measurements (SF), bio-electrical impedance analysis (BIA) and air displacement plethysmography (BOD-POD). RESULTS: No significant difference was found among measurements of FFM using the three different techniques. Both SenseWear and Sensor Medics Vmax showed a high correlation, r=0.42 and r=0.40 (p<0.0001) respectively, with BMI. No significant difference was found in mean REE between SenseWear (1540+/-280 kcal/day) and Sensor Medics Vmax (1700+/-330 kcal/day) (p=ns) and the correlation between REE measured by SenseWear and Sensor Medics Vmax was high (r=0.86, p<0.0001). Bland-Altman plot showed no difference in REE determination between SenseWear and Sensor Medics Vmax. ÿM determined by BOD-POD correlated with SenseWear (r=0.42, p<0.0001) as well as Sensor Medics Vmax (r=0.38, p<0.001). CONCLUSION: SF, BIA and BOD-POD provide valid and reliable measurements of FFM. Our results suggest that the SenseWear armband is an acceptable device to accurately measure REE in healthy subjects. Its characteristics have the potential to reduce measurement times and make the SenseWear armband useful for epidemiological studies.
2007
17
5
338
343
A new device for measuring resting energy expenditure (REE) in healthy subjects / Malavolti, Marcella; A., Pietrobelli; Dugoni, Manfredo; M., Poli; Romagnoli, Elisa; Battistini, Nino Carlo. - In: NMCD. NUTRITION METABOLISM AND CARDIOVASCULAR DISEASES. - ISSN 0939-4753. - STAMPA. - 17:5(2007), pp. 338-343. [10.1016/j.numecd.2005.12.009]
Malavolti, Marcella; A., Pietrobelli; Dugoni, Manfredo; M., Poli; Romagnoli, Elisa; Battistini, Nino Carlo
File in questo prodotto:
File Dimensione Formato  
NMCD-sensewearpaper.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 147.96 kB
Formato Adobe PDF
147.96 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/583739
Citazioni
  • ???jsp.display-item.citation.pmc??? 25
  • Scopus 132
  • ???jsp.display-item.citation.isi??? 114
social impact