Aims We investigated the applicability of the Ventricular Capture Control (VCC) and Atrial Capture Control (ACC) algorithms for automatic management of cardiac stimulation featured by Biotronik pacemakers in a broad, unselected population of pacemaker recipients. Methods and results Ventricular Capture Control and Atrial Capture Control were programmed to work at a maximum adapted output voltage as 4.8 V in consecutive recipients of Biotronik pacemakers. Ambulatory threshold measurements were made 1 and 12 months after pacemaker implant/replacement in all possible pacing/sensing configurations, and were compared with manual measurements. Among 542 patients aged 80 (73-85) years, 382 had a pacemaker implant and 160 a pacemaker replacement. Ventricular Capture Control could work at long term in 97% of patients irrespectively of pacing indication, lead type, and lead service life, performance being superior with discordant pacing/sensing configurations. Atrial Capture Control could work in 93% of patients at 4.8 V maximum adapted voltage and at any pulse width, regardless of pacing indication, lead type, and service life. At 12-month follow-up, a ventricular threshold increase ≥1.5 V had occurred in 4.4% of patients uneventfully owing to VCC functioning. Projected pacemaker longevity at 1 month was strongly correlated with the 12-month estimate, and exceeded 13 years in >60% of patients. Conclusion These algorithms for automatic management of pacing output ensure patient safety in the event of a huge increase of pacing threshold, while enabling maximization of battery longevity. Their applicability is quite broad in an unselected pacemaker population irrespectively of lead choice and service of life.

Automatic management of atrial and ventricular stimulation in a contemporary unselected population of pacemaker recipients: The ESSENTIAL Registry / Biffi, Mauro; Bertini, Matteo; Saporito, Davide; Belotti, Giuseppina; Quartieri, Fabio; Piancastelli, Maurizio; Pucci, Angelo; Boggian, Giulio; Mazzocca, Gian Franco; Giorgi, Davide; Diotallevi, Paolo; Diemberger, Igor; Martignani, Cristian; Pancaldi, Stefano; Ziacchi, Matteo; Marcantoni, Lina; Toselli, Tiziano; Attala, Simone; Iori, Matteo; Bottoni, Nicola; Argnani, Selina; Tomasi, Corrado; Sassone, Biagio; Boriani, Giuseppe. - In: EUROPACE. - ISSN 1099-5129. - 18:10(2016), pp. 1551-1560. [10.1093/europace/euw021]

Automatic management of atrial and ventricular stimulation in a contemporary unselected population of pacemaker recipients: The ESSENTIAL Registry

BORIANI, Giuseppe
2016

Abstract

Aims We investigated the applicability of the Ventricular Capture Control (VCC) and Atrial Capture Control (ACC) algorithms for automatic management of cardiac stimulation featured by Biotronik pacemakers in a broad, unselected population of pacemaker recipients. Methods and results Ventricular Capture Control and Atrial Capture Control were programmed to work at a maximum adapted output voltage as 4.8 V in consecutive recipients of Biotronik pacemakers. Ambulatory threshold measurements were made 1 and 12 months after pacemaker implant/replacement in all possible pacing/sensing configurations, and were compared with manual measurements. Among 542 patients aged 80 (73-85) years, 382 had a pacemaker implant and 160 a pacemaker replacement. Ventricular Capture Control could work at long term in 97% of patients irrespectively of pacing indication, lead type, and lead service life, performance being superior with discordant pacing/sensing configurations. Atrial Capture Control could work in 93% of patients at 4.8 V maximum adapted voltage and at any pulse width, regardless of pacing indication, lead type, and service life. At 12-month follow-up, a ventricular threshold increase ≥1.5 V had occurred in 4.4% of patients uneventfully owing to VCC functioning. Projected pacemaker longevity at 1 month was strongly correlated with the 12-month estimate, and exceeded 13 years in >60% of patients. Conclusion These algorithms for automatic management of pacing output ensure patient safety in the event of a huge increase of pacing threshold, while enabling maximization of battery longevity. Their applicability is quite broad in an unselected pacemaker population irrespectively of lead choice and service of life.
2016
14-giu-2016
18
10
1551
1560
Automatic management of atrial and ventricular stimulation in a contemporary unselected population of pacemaker recipients: The ESSENTIAL Registry / Biffi, Mauro; Bertini, Matteo; Saporito, Davide; Belotti, Giuseppina; Quartieri, Fabio; Piancastelli, Maurizio; Pucci, Angelo; Boggian, Giulio; Mazzocca, Gian Franco; Giorgi, Davide; Diotallevi, Paolo; Diemberger, Igor; Martignani, Cristian; Pancaldi, Stefano; Ziacchi, Matteo; Marcantoni, Lina; Toselli, Tiziano; Attala, Simone; Iori, Matteo; Bottoni, Nicola; Argnani, Selina; Tomasi, Corrado; Sassone, Biagio; Boriani, Giuseppe. - In: EUROPACE. - ISSN 1099-5129. - 18:10(2016), pp. 1551-1560. [10.1093/europace/euw021]
Biffi, Mauro; Bertini, Matteo; Saporito, Davide; Belotti, Giuseppina; Quartieri, Fabio; Piancastelli, Maurizio; Pucci, Angelo; Boggian, Giulio; Mazzocca, Gian Franco; Giorgi, Davide; Diotallevi, Paolo; Diemberger, Igor; Martignani, Cristian; Pancaldi, Stefano; Ziacchi, Matteo; Marcantoni, Lina; Toselli, Tiziano; Attala, Simone; Iori, Matteo; Bottoni, Nicola; Argnani, Selina; Tomasi, Corrado; Sassone, Biagio; Boriani, Giuseppe
File in questo prodotto:
File Dimensione Formato  
euw021.pdf

Accesso riservato

Tipologia: Versione pubblicata dall'editore
Dimensione 383.25 kB
Formato Adobe PDF
383.25 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

Licenza Creative Commons
I metadati presenti in IRIS UNIMORE sono rilasciati con licenza Creative Commons CC0 1.0 Universal, mentre i file delle pubblicazioni sono rilasciati con licenza Attribuzione 4.0 Internazionale (CC BY 4.0), salvo diversa indicazione.
In caso di violazione di copyright, contattare Supporto Iris

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11380/1139538
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 12
social impact